EXTERNAL PUMP CONTROL
This section of the PHD Ultra manual describes the control of the pump using an external computer device.
Setting Up HyperTerminal
HyperTerminal is a Windows application designed to support the external control of devices such as the
PHD Ultra pump through an RS-232 or USB connection. The following instructions describe the
configuration of the HyperTerminal application.
1. Select “Start – All Programs – Accessories – Communications – HyperTerminal”.
2. Enter a name for a New Connection (i.e PHD_Ultra_Comm), then click OK.
3. Select the Virtual Comm Port from the “Connect using” drop-down list. Click OK.
(If the Virtual Comm Port is not known, use Device Manager to find it. Instructions on setting up the
Virtual CommPort Driver are supplied in Appendix E.)
4. Set up the Port Settings as shown above and click OK.
Harvard Apparatus PHD ULTRA
Syringe Pump Series User's Manual
Publication 5419-002-Rev-A
81
U
sing The PHD ULTRA™