User`s guide
AU TOFILL PROF ILE
The Autofill flow profile is an automated infuse/withdraw or withdraw/infuse step that operates at
constant, but separately defined, flow rates for the withdraw and infuse operations. The Autofill profile is
designed to work in conjunction with an autofill valve that controls the switching of flows from a reservoir
for refilling the syringe, to the target infusion subject.
Figure 63: Autofill Flow Pump Operation Profile.
When creating or editing an Autofill Rate profile, you will define the following operating parameters:
• The Flow Mode defining the starting direction of flow: infuse or withdraw
• The Flow Rates to be used during infuse/withdraw operations
• The infusion volume to be delivered during each pump cycle
• The total volume to be delivered during all cycles or the total number of infuse/withdraw cycles that
you want the pump to execute. For example, if you want to pump 100 ml using a 10 ml syringe, the
pump will execute 10 Infuse/Withdraw cycles
During an Autofill profile, the PHD ULTRA
™
operates at a constant pump rate during the infusion and
withdrawal stages, although the rate can be different for each direction (i.e. infuse vs. withdraw).
When creating or editing an Autofill flow profile, you can set the rate in ml, µl, nl or pl units of volume
using sec, min or hr as the units of time. The system automatically calculates a minimum and maximum
flow rate based on the syringe selected or custom-defined. The min/max values are made available as
button selection options on the Set Rate and Concentration screen.
Figure 64: Step Definition screen for an Autofill profile.
✐
NOTE: Pin 8 of the 15-pin I/O Connector adapter on the rear of the PHD ULTRA™ will
actuate to control a Harvard Apparatus Valve Box used in conjunction with an Autofill Pump
Control Profile.
Flow
Rate
Withdraw
Infuse
Time | Volume
Harvard Apparatus PHD ULTRA
™
Syringe Pump Series User's Manual
Publication 5419-002-Rev-A
66
U
sing The PHD ULTRA™