4. On the Units Selection screen, select the unit of measure (ml, µl, nl or pl) and the unit of time (sec,
min, hr) as appropriate, then choose Accept to confirm your selections.
5. Select either the Volume or Time button to determine how you will control the Bolus infusion. If you
selected Time, continue with Step 6. If you selected Volume, proceed to Step 9.
6
. On the Set Target Times screen, enter the desired value using the keypad.
7. If the Units of Time are not correct for your application, select the Units button to toggle between sec
(seconds) and h:m:s (hours:minutes:seconds). To use the h:m:s option, enter your time value in hours,
minutes and seconds, separating each value with a colon.
8. When the Time parameters are correctly set, choose the Accept button to return to the Step Definition
screen.
9. On the Set Target Volumes screen, enter the desired value using the keypad.
10. If the Units of Measure are not correct for your application, choose the Units button to enter the Units
selection screen. Select the unit of measure button (ml, µl, nl, pl) applicable for your application, then
choose the Accept button to return to the Set Target Volumes screen.
11. When the Volume parameters are correctly set, choose the Accept button to return to the Step
Definition screen.
Harvard Apparatus PHD ULTRA
™
Syringe Pump Series User's Manual
Publication 5419-002-Rev-A
62
U
sing The PHD ULTRA™