The Ramp Rate profile is used to create a linearly increasing or decreasing flow rate for a single pumping
mode (infuse or withdraw) over the course of a defined time period.
Figure 48: Ramp Rate Pump Operation Profile.
When creating or editing a Ramp Rate profile, you will define three primary operating parameters:
The Flow Mode defining the direction of flow: infuse or withdrawl
The Flow Rate at the beginning and end of the pumping cycle
The Target Time that determines the total length of time over which the ramp will be calculated
During a Ramp Rate profile, the PHD ULTRA
varies the flow rate linearly between the starting rate and
ending rate over the full time period entered. A Ramp Rate profile can be applied to an increasing or
decreasing rate application.
When creating or editing a Ramp Rate profile, you can set the rate in ml, µl, nl or pl units of volume using
sec or h:m:s (hours:minutes:seconds) as the units of time. The system automatically calculates a minimum
and maximum flow rate based on the syringe selected or custom-defined. The min/max values are made
available as button selection options on the Set Infuse and Withdraw Rate screens.
TO SET UP/EDIT A RAMP RATE PROFILE:
1. From the Step Definition screen for a Ramp Rate profile, choose the Set Mode button to toggle the
Flow Mode between Infuse and Withdraw operation.
Figure 49: Step Definition screen for a Ramp Rate profile.
2. Choose the Set Start & End Rate button to enter the Set Start & End Rates screen.
Flow
Rate
Time
S
TOP
T
arget Time
Harvard Apparatus PHD ULTRA
Syringe Pump Series User's Manual
Publication 5419-002-Rev-A
55
U
sing The PHD ULTRA™