User`s guide
4. When you have entered the desired name for your Method, choose the Accept button to return to the
Method Main screen. Your new Method will now be listed in the Method Select button.
5. Choose the Syringe Select button and complete the description of the syringe by following the
procedure outlined previously in the Syringe Selection section of this Guide.
6
. To begin creating Pump Control Profile Steps, choose the Step Definition button to enter the Method
Steps Summary screen.
7. To add a Step to your Method, choose the Add Step button to access the Method Steps Menu screen.
The available options include:
Figure 43: Method Steps Profile Menu screen.
8. Choose the Method Step type that you want to add to the Method. If you choose a pumping profile
(e.g. Constant Rate, Ramp, etc.) you will be taken to the Step Definition screen for the selected
Method. If you choose Advanced Options, you will taken to the Select Advanced Options screen.
9. Complete the parameter entry applicable to the selected Pump Operation Profile, follow the detailed
instructions supplied in the Pump Operation Profiles (Preloaded and User-Defined Methods) section
later in this Guide. If you selected Advanced Options, follow the instructions supplied in the
Incorporating Advanced Options Into a Method section later in this manual.
10. When you have completed all parameter entries on the Step Definition screen or completed the
required entries for Advanced Options, choose the Accept button to save your changes and return to
the Method Steps Summary screen.
11. Repeat items 7-10 in this procedure to include additional Steps in the Method.
12. When you are done adding all required Steps to the Method, choose the Back button to return to the
Method Main screen.
✐
NOTE: When adding a Step to a Method, the Step is added after the currently selected
Step (i.e. the Step that is highlighted in purple)
✐
NOTE: If you do not complete the required parameter entry for a Step, subsequent
Step additions may be ignored. You should enter all required parameters for a Step
when creating it to ensure that the application software can assemble your Method
without errors.
c. Stepped
d. Pulse
e. Bolus
f. Concentration
a. Constant Rate
b. Ramp
g. Autofill
h. Advanced Options
Harvard Apparatus PHD ULTRA
™
Syringe Pump Series User's Manual
Publication 5419-002-Rev-A
49
U
sing The PHD ULTRA™