User`s guide
2. Choose the Syringe Select button to enter the Syringe Selection screen.
3. Follow the instructions in the Syringe Selection section of this Guide to complete the selection of a
syringe from the built-in syringe definition list or to create a new syringe definition. When finished,
choose the Accept button to save your changes, or choose the Cancel button to cancel your changes,
a
nd return to the Quick Start screen.
4. Choose the Infuse Rate Select button to access the Set Infuse and Withdraw Rates screen.
5. Enter the required parameters on the Set Infuse and Withdraw Rates screen. Details on setting the
Infuse and Withdraw Rates are provided in the Quick Start Method Variables section of this Guide.
When finished, choose the Accept button to save your changes, or choose the Cancel button to cancel
your changes, and return to the Quick Start screen.
6. Choose the Target Volume Select button to enter the Set Target Volume screen.
7. Enter the required data on the Set Target Volume screen as described in the Quick Start Method
Variables section of this Guide.
Quick Start Method Variables
IN FUSE RATE S ELEC T/W ITHD RAW RATE SELECT
The Quick Start Infuse Rate and Withdraw Rate Select functions define the rate of pumping while infusing
or withdrawing liquids respectively. The two functions are controlled identically, so we have described the
operation here only once.
When adjusting the Infuse Rate, you can set the rate in ml, µl, nl or pl units of volume and use sec, min or
hr as the units of time. The system automatically calculates a minimum and maximum flow rate based on
the syringe selected or custom-defined. The min/max values are made available as button selection options
on the Set Infuse and Withdraw Rate screens.
Figure 35: Set Infuse Rate screen in a Quick Start Infuse Only Method.
✐
NOTE: The following instructions describe setting the Infuse Rate. The instructions are
the same for Withdraw Rate. On an Infuse/Withdraw combination Method, you can set
both rates on a single screen by toggling between the two rate options using the Infuse
Rate or Withdraw Rate buttons at the top of the screen.
Harvard Apparatus PHD ULTRA
™
Syringe Pump Series User's Manual
Publication 5419-002-Rev-A
42