User`s guide

Figure 23: Method Main screen showing a sample Method with 10 steps.
SY RINGE S ELEC TIO N
In order to accurately control the flow rate or volume delivery of the pump, the PHD ULTRA
requires that
each method include accurate setting of the syringe diameter and volume. All Methods (Quick Start,
Preloaded, User-Defined) require a syringe definition, either by selecting from the installed syringe
specifications list or by defining a new (Custom) syringe in the application.
The built-in syringe specifications list provides a convenient lookup function that allows users to select
from a wide variety of syringes from major manufacturers, including: Air-Tite, Becton Dickinson, Hamilton,
Harvard Apparatus, Cadence Science, SGE Scientific Glass Engineering, Sherwood – Monoject, and Terumo.
This Guide also includes a reference table of major syringe diameter and volume parameters in Appendix A:
Syringe Volume/Diameter Reference Table.
The following parameters must be set for every syringe selection:
Syringe Diameter: mm
Syringe Volume: ml or µl
TO SELECT A SYRINGE USING THE SYRINGE LOOKUP FUNCTION:
1. From the Quick Start or Method Main screen, choose the Syringe Select button.
Figure 24: Syringe Selection screen with currently selected syringe.
!
WARNING: When using low-volume, small diameter syringes or pumping
high-viscosity liquids, you should adjust the Force value in the Settings
menu to avoid accidental damage to the syringe.
Harvard Apparatus PHD ULTRA
Syringe Pump Series User's Manual
Publication 5419-002-Rev-A
35
U
sing The PHD ULTRA™