SYSTEM SETTING ADJUSTMENTS
The PHD ULTRA
™
Settings menu is used to control basic system operations. This section describes the
options available in this section of the application software.
.
Figure 8: The Settings screen.
Setting the Audible Alarms
The PHD ULTRA
™
pump incorporates an Audible Alarm feature that can be used to warn the operator
about various operating conditions, errors and power outages. The following alarms are user configurable
in the Settings section of the application:
• End-of-Run Notice: An alarm will sound when a Method run is completed.
• Power Up: An alarm will sound if power to the PHD ULTRA
™
is lost while the pump is running.
Pressing any button on the screen silences this alarm.
• Motor Stall: Notifies the operator of any motor stall.
• Keypad Clicks: A soft beep indicates a positive touch on any button on the screen.
• Near End-of-Run: This alarm can be custom configured to notify the operator that a Method is near-
completion. The user can select from preset values of 75, 80, 85, 90 or 95% completion, or can set a
custom value using the touchscreen.
• Calibration Reminder: This alarm is used to inform the user that the system has reached its scheduled
Calibration Update date.
• Mute All: Selecting this option mutes all audible alarms. When selected, the Mute All button is
indicated with a checkmark and the remaining buttons on the screen are greyed out.
Figure 9: Set Audible Alarms screen showing the available options and with Mute All selected. The remaining
alarm options are greyed out in this example.