User`s guide
Upgrading EZ PRO Software
1. Upload the latest software version to your desktop (format is filename.dfu).
2. Connect the pump to PC using a USB cable. Open HyperTerminal and verify communication with the
p
ump.
3. Type “boot” into HyperTerminal. The pump’s screen will be blank and the light should be flashing.
4. Disconnect then reconnect the USB cable as instructed by HyperTerminal.
5. From the CD provided with the pump, open the PHD ULTRA
™
Flash Upgrade Application.
6. Check “Verify after download” then click browse. Browse to the file previously saved on your desktop
and click open.
7. Click upgrade. The previous software will first be erased, the new software will be loaded, and finally
the software will be verified.
8. Disconnect the USB cable and power cycle the pump.
9. Verify that the version number has been updated in the Power Up screen.
Version Number
✐
N
OTE: First time users will need to install the PHD ULTRA™ USB driver as well as
the bootloader driver file (PHDUltra-Bootloader.inf). Reference “USB Virtual Comport
Driver Installation”.
Harvard Apparatus PHD ULTRA
™
Syringe Pump Series User's Manual
Publication 5419-002-Rev-A
112