Technical data

94 Agilent 1290 Infinity LC System Manual and Quick Reference
5 Quick Start Guide
Preparing the system
3 In the Available Signals box, highlight the required signal(s), and click Add to
move them to the Selected Signals box.
4 To configure the individual settings for each signal, highlight the signal in
the Selected Signal box and set the required values in the lower half of the
page.
The Online Plot window behaves like electronic chart paper, continuously
recording the output from the detector(s) and other output parameters. The
signals are drawn at the right of the window and move away to the left. Up to
90 min of past data is accessible. This is useful for checking the baseline and
looking at previous injections. The X and Y-axis scales can be adjusted directly
with the up/down buttons on each axis.
The Adjust button in the Online Plot window moves the current point on the
selected signal to the zero line. The selected signal is indicated by the color of
the Y-axis labels. A particular signal may be selected by clicking on the signal
or by clicking on the relevant signal description at the top of the plot.
The Balance button zeroes all detectors when pressed.
NOTE
In addition to the detector signals, parameter traces such as temperature and pressure can
also be plotted. With Apply to Method, the settings in this page can be stored into the
method.
NOTE
Changes made in the Online Plot page do not in any way affect the data stored into the
individual data files.