Technical data
Agilent 1290 Infinity LC System Manual and Quick Reference 81
System Setup and Installation
4
Installing the Module
Two Stack Configuration
In case the autosampler thermostat is added to the system, a two-stack
configuration is recommended, which places both heavy modules (1290
Infinity pump and thermostat) at the bottom of each stack and avoids high
stacks. Some users prefer the lower height of this arrangement even without
the autosampler thermostat. A slightly longer capillary is required between
the pump and autosampler. (See Figure 31 on page 81 and Figure 32 on
page 82).
Figure 31 Recommended two stack configuration for 1290 Infinity with quaternary pump
(front view)
Instant Pilot
Detector
Column compartment
Solvent cabinet
Pump
Autosampler
Thermostat for the
ALS (optional)