Technical data
Agilent 1290 Infinity LC System Manual and Quick Reference 59
Optimization of the Agilent 1290 Infinity LC System
3
How to Achieve Higher Sensitivity
How to Achieve Higher Sensitivity
The sensitivity of a separation method is linked to the choice of stationary and 
mobile phases as good separation with narrow peaks and a stable baseline 
with minimal noise are desirable. The choice of instrument configuration will 
have an effect and a major impact is the setup of the detector. This section 
considers how sensitivity is affected by:
• Pump mixer volume
• Narrower columns
• Detector flow cell
• Detector parameters
In addition, the discussion on detector parameters also mentions the related 
topics of selectivity and linearity.
Pump Mixer Volume
To achieve lowest baseline noise with UV detection, the standard delay volume 
configuration with the 35 µl Jet Weaver is recommended for the 1290 Infinity 
pump module. This is applicable for almost all applications but when using 
TFA in the mobile phase, or other situations requiring more mixing, the larger 
volume side of the Jet Weaver mixer should be used to minimize mixing noise. 
Columns
Sensitivity is specified as a signal-to-noise ratio (S/N) and hence the need to 
maximize peak height and minimize baseline noise. Any reduction in peak 
dispersion will help to maintain peak height and so extra-column volume 
should be minimized by use of short, narrow internal diameter, connection 
capillaries and correctly installed fittings. Using smaller inner diameter 
columns should result in higher peak height and is therefore ideal for 
applications with limited sample amounts. If the same sample amount can be 
injected on a smaller i.d. column, then the dilution due to column diameter 
will be less and the sensitivity will increase. For example, decreasing the 
column i.d. from 4.6 mm to 2.1 mm results in a theoretical gain in peak height 
of 4.7 times due to the decreased dilution in the column. For a mass 
spectrometer detector, the lower flow rates of narrow columns can result in 
higher ionization efficiencies and therefore higher sensitivity.










