Technical data
54 Agilent 1290 Infinity LC System Manual and Quick Reference
3 Optimization of the Agilent 1290 Infinity LC System
How to Achieve High Throughput
The injection can be optimized for speed remembering that drawing the 
sample too fast can reduce the reproducibility. Marginal gains are to be made 
here as the sample volumes used tend towards the smaller end of the range in 
any case. A significant portion of the injection time is the time taken with the 
needle movements to and from the vial and into the flush port. These 
manipulations can be performed while the previous separation is running. 
This is known as "overlapped injection" and it can be easily turned on from the 
autosampler setup screen in the ChemStation Control Software. The 
autosampler can be told to switch the flow through the autosampler to bypass 
after the injection has been made and then after, for example, 3 minutes into a 
4 minutes run to start the process of aspirating the next sample and preparing 
for injection. This can typically save 0.5 to 1 minute per injection. For sticky 
compounds it is recommend to do this during column equilibration, when the 
autosampler has seen the starting conditions of the next gradient run.
The column equilibration step can be a significant part of the cycle time. 
Typically the column needs flushing with three to five times the column 
volume to stabilize it ready for the next injection and this can be 50 % or more 
of the separation time in some applications. It is an essential process but can 
be taken out of the cycle time by using automated alternating column 
regeneration. For this a two-position ten-port, 1200 bar, valve head is required 
in the column compartment; a second analytical column, identical to the first; 
and a second pump is needed. As one column is being used in the separation 
run, the other column is being flushed with the starting composition of the 
mobile phase gradient and to start the next injection the newly re-equilibrated 
column is switched into the analytical flow path. The two columns then 
alternate in this way for the whole sequence of injections. The second pump is 
only required to flush an isocratic mixture through the column and so can be a 
simpler pump than the 1290 Infinity pumps. For instance a 1200 Series 
isocratic pump would be sufficient to perform this task. The setup is 
illustrated in Figure 22 on page 55.










