Technical data
Agilent 1290 Infinity LC System Manual and Quick Reference 51
Optimization of the Agilent 1290 Infinity LC System
3
How to Achieve Higher Injection Volumes
How to Achieve Higher Injection Volumes
The standard configuration of the Agilent 1290 Infinity Autosampler includes 
a variable volume sample loop for up to 20 µl injections. The metering device 
can inject a maximum volume of 40 µl and the sample loop cartridge can be 
exchanged to allow this (refer to the 1290 Infinity Autosampler manual for 
details). The system delay volume due to the autosampler will increase 
accordingly. 
Whenever a method is scaled down from a larger column to a smaller column 
it is important that the method translation makes an allowance for reducing 
the injection volume in proportion to the volume of the column to maintain 
the performance of the method. This is to keep the volume of the injection at 
the same percentage volume with respect to the column. This is particular 
important if the injection solvent is stronger (more eluotropic) than the 
starting mobile phase and any increase will affect the separation particularly 
for early running peaks (low retention factor). In some cases it is the cause of 
peak distortion and the general rule is to keep the injection solvent the same 
or weaker than the starting gradient composition. This has a bearing on 
whether, or by how much, the injection volume can be increased and the user 
should check for signs of increased dispersion (wider or more skewed peaks 
and reduced peak resolution) in trying to increase the injection size. If an 
injection is made in a weak solvent then the volume can probably be increased 
further because the effect will be to concentrate the analyte on the head of the 
column at the start of the gradient. Conversely if the injection is in a stronger 
solvent than the starting mobile phase then increased injection volume will 
spread the band of analyte down the column ahead of the gradient resulting in 
peak dispersion and loss of resolution. 
Perhaps the main consideration in determining injection volume is the 
diameter of the column as this will have a big impact on peak dispersion. Peak 
heights can be higher on a narrow column than with a larger injection on a 
wider column because there is less peak dispersion. With 2.1 mm i.d. columns 
typical injection volumes might range up to 5 to10 µl but it is very dependent 
on the chemistry of the analyte and mobile phase as discussed above. A rough 
guide to maximum injection volume can be gained by looking at the volume of 
the column (see Table 7 on page 44) – in a gradient separation injection 
volumes of about 5 % of the column volume might be achieved whilst 
maintaining good resolution and peak dispersion. 










