Technical data
Agilent 1290 Infinity LC System Manual and Quick Reference 49
Optimization of the Agilent 1290 Infinity LC System
3
How to Configure the Optimum Delay Volume
Installing the V380 Jet Weaver in the 1290 Infinity Quaternary Pump
1 Open the screw of the Jet Weaver metal lid. 2 Remove the metal lid by lifting it up (1) and pulling it out of 
the front panel (2).
3 Insert the Jet Weaver to the opening in the front panel (1) 
and push it down (2). 
4 Mount both capillary connections to the Jet Weaver 
observing the correct orientation.
1
2
1
2










