Technical data
42 Agilent 1290 Infinity LC System Manual and Quick Reference
3 Optimization of the Agilent 1290 Infinity LC System
How to Configure the Optimum Delay Volume
How to Configure the Optimum Delay Volume
Table 3 on page 42 and Table 4 on page 43 show the component volumes which 
contribute to system delay volume in the Agilent 1290 Infinity LC System. In 
the standard configuration with the Agilent 1290 Infinity Binary Pump, with 
the Jet Weaver mixer, the 1290 Infinity Autosampler and Thermostatted 
Column Compartment the system delay volume is about 125 µl. This standard 
delay volume is suitable for the majority of applications. For example, a fast 
separation on a 50 mm x 2.1 mm column packed with sub-two-micron 
particles at a moderate flow rate of 0.6 ml/min will result in a typical gradient 
delay time of about 0.2 min which is generally acceptable with gradient times 
of two to three minutes (see Table 6 on page 43). It is often useful to consider 
the flow rate in terms of column volumes and it can be seen from Table 7 on 
page 44 that with this column at 0.6 ml/min about 6 column volumes per 
minute flow through the system and the delay volume is about 1.2 times the 
column volume.
A setup with the Agilent 1290 Infinity Quaternary Pump, the 1290 Infinity 
Autosampler, and Thermostatted Column Compartment has a delay volume of 
430 µL, resulting in a delay time of 0.7 min. This is the limit for acceptance of 
gradient times of 3 min.
Tabl e 3 Delay volumes of 1290 Infinity LC modules
Components Delay Volume (µl)
Binary Pump  10
Jet Weaver mixer (standard) 35
Binary Pump + Jet Weaver 45
Quaternary Pump 350
Quaternary Pump + V380 Jet Weaver  500
Autosampler (fixed 5 µl loop) 5
Autosampler (standard, variable volume) 80
Column Compartment low-dispersion heat exchanger 1.6
Connection tubing, 0.12 mm i.d., per 100 mm 1.1










