Technical data
Agilent 1290 Infinity LC System Manual and Quick Reference 41
Optimization of the Agilent 1290 Infinity LC System
3
Delay Volume and Extra-Column Volume
times of about one minute. However, the delay volume must be reduced further 
to achieve delay times which are a fraction of the intended run time. This is 
achieved with the Agilent 1290 Infinity LC System due to the low delay volume 
of the pump flow path, low volume of the Jet Weaver mixer and low-volume of 
the flow path through the autosampler.
Extra-Column Volume
Extra-column volume is a source of peak dispersion that will reduce the 
resolution of the separation and so should be minimized. Smaller diameter 
columns require proportionally smaller extra-column volumes to keep peak 
dispersion at a minimum. 
In a liquid chromatograph the extra-column volume will depend on the 
connection tubing between the autosampler, column and detector; and on the 
volume of the flow cell in the detector. The extra-column volume is minimized 
with the Agilent 1290 Infinity LC System due to the narrow-bore (0.12 mm 
i.d.) tubing, the low-volume heat exchangers in the column compartment and 
the Max-Light cartridge cell in the detector. 