Technical data
Agilent 1290 Infinity LC System Manual and Quick Reference 35
The Agilent 1290 Infinity LC System - Product Description
2
System Components
Figure 19 Quick change valve in TCC
Up to three TCC can be “clustered” to allow advanced applications such as 
switching between eight columns for automated method development or to 
make additional columns available for different applications. Thus, the column 
to be used becomes a simple method parameter. This requires two 8 position/9 
port valve heads, one each in two of the TCCs. Clustered TCC are represented 
by the software as one unit with one interface for ease of operation. 
Further improvements compared to earlier designs include better thermal 
insulation, better capillary guides and a “door open” sensor so that methods 
can define that the door must be closed – especially useful for low or 
high-temperature methods. 