Technical data
Agilent 1290 Infinity LC System Manual and Quick Reference 33
The Agilent 1290 Infinity LC System - Product Description
2
System Components
A totally new optional add-on module, the Flexible Cube, works seamlessly 
with the autosampler to provide additional capabilities. With the addition of 
the new Flexible Cube module, which comprises a 500 µl syringe pump, one 
low pressure valve and two high pressure switching valves further options 
become possible. For instance, the flow-through injection system can be 
replaced by a fixed injection loop using the syringe pump and valving of the 
Flexible Cube to fill the sample loop. The advantage of this configuration is 
that it eliminates the delay volume from the autosampler and so may be 
preferred in some fast-gradient high-throughput situations. The trade-off is 
that the flexibility of variable volume injection is turned off and some sample 
is wasted in flushing the loop. Other tasks such as automatically back-flushing 
the injection needle seat after injection are possible with the Flexible Cube 
giving further confidence that carryover of difficult compounds or blockages 
caused by dirty samples can be routinely avoided.
The autosampler sample rack has 10 fixed positions for 2 ml vials and two 
removable trays, which may or may not be identical, selected from: 
• Tray for 2 ml vials with 54 positions 
• Microtitre plate with 96 well positions (various heights configurable)
• Microtitre plate with 384 well positions (various heights configurable)
If required the autosampler can be thermostatted from 4 °C to 40 °C by the 
addition of the autosampler temperature control module.