Technical data
Agilent 1290 Infinity LC System Manual and Quick Reference 27
The Agilent 1290 Infinity LC System - Product Description
2
System Components
chromatographic performance these features makes the pump very quiet in 
operation.
When concentrated buffer solutions are used as mobile phase, the active seal 
wash option is available for use to extend the life of the pump seals.
A solvent selection valve allows binary mixtures (isocratic or gradient) to be 
formed from one of two solvents per channel. Binary gradients are created at 
the purge valve by high-pressure mixing of solvents from pump A and pump B. 
The purge valve allows the flow to be switched under software control to 
waste for purging new solvents through the pump head. A sensor is connected 
to the purge valve to monitor the system pressure.
Figure 13 The Jet Weaver mixer