Technical data
26 Agilent 1290 Infinity LC System Manual and Quick Reference
2 The Agilent 1290 Infinity LC System - Product Description
System Components
Figure 12 Parts identification and schematic of the 1290 Infinity Binary Pump
Each pump head is a dual piston in-series design utilizing novel firmware 
control and novel piston material, silicon carbide, which efficiently removes 
heat from the pump. The capillary joining the primary piston to the secondary 
piston has an integrated heat exchanger to remove the heat created at high 
pressure and flow. Each pump head has one passive inlet valve and one 
passive outlet valve on the primary piston chamber. Each piston is 
independently and precisely driven by a motor with 65000 steps giving 300 
picolitres displacement per step.
The movement of the pistons is under intelligent control with a feedback loop 
to ensure that this active damping of pressure pulsation results in a ripple-free 
flow. The piston drive tunes itself for the compressibility characteristics of the 
solvent and the hydraulic characteristics of the system to maintain the 
ripple-free state. This in conjunction with the smooth-motion control, which 
reduces pressure pulsation caused by the movement of the piston, and the 
low-volume efficient mixing ensure that pump noise on UV traces is the lowest 
possible. A dedicated microprocessor in the pump takes care of the smooth 
motion control and optimization of the pistons’ motion for real-time 
optimization based on the static and dynamic parameters. In addition to the 
;gdbhdakZcigZhZgkd^gh
EgZhhjgZ
hZchdg
IdVjidhVbeaZg
B^mZg
Eg^bVgn
ejbe]ZVY
Eg^bVgn
ejbe]ZVY
HZXdcYVgn
ejbe]ZVY
HZXdcYVgn
ejbe]ZVY
9Z\VhhZg
IdlVhiZ
8]VccZa7
8]VccZa6
HdakZci
hl^iX]^c\
kVakZ