Technical data
18 Agilent 1290 Infinity LC System Manual and Quick Reference
1 Introduction to Ultra-High Performance Liquid Chromatography
Frictional Heating
Frictional Heating
Forcing mobile phase through the column at higher pressure and higher flow 
rates generates heat. The resulting temperature gradients (radial and 
longitudinal) can have an impact on the column efficiency.
where F is the flow rate and p is the pressure.
Powerful column thermostatting (for example, using a water bath) generates a 
strong radial temperature gradient, which leads to significant loss in column 
efficiency. Still-air column thermostatting reduces the radial temperature 
gradient and therefore reduces the efficiency losses, but a higher column 
outlet temperature has to be accepted. The raised temperature may have an 
effect on selectivity. At lower back-pressure, performance losses due to 
frictional heat are minimized so that 4.6 or 3 mm inner diameter sub-2-micron 
columns still deliver superior efficiencies compared with the respective 
2.1 mm inner diameter columns.
An example of a gradient method transfer to a 2.1 mm i.d. STM column in 
which the separation has been speeded up is shown in Figure 9 on page 19. 
The initial run on the 2.1 mm column was at a flow rate of 0.22 ml/min 
generating 380 bar pressure at a temperature setting of 37 °C with all peaks 
separated in 12.5 min (chromatogram not shown). The flow was increased to 
0.66 ml/min and the gradient times adjusted downwards by a factor of three 
generating 1020 bar pressure with all peaks eluted in 4.2 min (Figure 9 on 
page 19 above). This should give the same separation but a loss of resolution 
can be seen between peaks 7 and 8 and between 5 and the main peak and this 
is due to heating in the column changing the selectivity for these compounds. 
It was found that running the column thermostat at 5 °C lower was sufficient 
to offset the heating effect inside the column and restore the separation 
(Figure 9 on page 19 below). The pressure increased to 1070 bar which also 
indicates that the in-column temperature was lower.










