Technical data
136 Agilent 1290 Infinity LC System Manual and Quick Reference
6 Appendix
Setting Up a Method using Edit Entire Method
The Edit Integration Events screen has two tables:
• Initial Events For All Signals contains events (integration parameters) that 
apply to all signals acquired with the method,
• Specific Events For Signal contains events which are specific for one type of 
detector or specific to different signals from the same detector. 
Key parameters in this table are:
• Slope Sensitivity represents the slope and curvature of the baseline 
needed to mark the start and end of a peak;
• Peak Width: the width at half-height of the narrowest peak of interest 
should be entered. This helps the integrator to distinguish between noise 
and very small peaks;
• Area Reject / Height Reject values that control rejection from the results of 
peaks whose area or height falls below these values;
• Integration OFF/ON suppresses integration between set limits. Almost 
always used to inhibit integration in the region from injection to solvent 
front or unretained peak marker.
Lines such as Integration OFF/ON are added to the table using the icons at the 
top of the window.
Click OK to exit and the next screen in the Edit Entire Method process will 
open.
Specify Report 
The Specify Report screen can also be directly accessed in Method and Run Control 
view by right-clicking on the graphical user interface on the Report icon and 
then selecting Specify Report in the context menu. In the Data Analysis view it 
can be accessed through the menu Report > Specify Report or the Specify Report 
task icon.