Technical data
130 Agilent 1290 Infinity LC System Manual and Quick Reference
6 Appendix
Setting Up a Method using Edit Entire Method
• Store controls the spectral collection mode with the following options:
None – no spectra stored,
Apex+Baselines – 3 spectra taken at start, apex and end of peak,
Apex+Slopes+Baselines – 5 spectra taken at start, upslope, apex, 
downslope and end of peak,
All in Peak – all the available spectra within a peak are stored,
All – all spectra throughout the run are stored,
Every 2nd Spectrum – stores only alternate spectra acquired throughout 
the run.
• Range – Spectra can be saved across the full range of the detector, 190 nm 
to 640 nm, or any reduced range chosen as appropriate by the user. (This 
also reduces the number of data to be stored). 
• Step – controls the interval (nm) of the data stored in a spectrum and 
hence affects the spectral resolution seen. The default setting of 2 nm is a 
good choice for most applications.
• Threshold – determines that spectra are not stored for peak heights below 
this value (mAU).
• Advanced - Analog Output
The 1290 Infinity detector has one analog signal output connector for use 
with data systems that do not accept a digital input. The following may be 
set:
• Zero Offset places the zero level at a set percentage of the output signal 
hence allowing some scope for negative drift,
• Attenuation scales the set absorbance to the full output.
• Advanced - Margin for Negative Absorbance
The default setting is 100 mAU, which means that the detector has 
sufficient dynamic range, considering where the zero level was set, to 
measure down to this value. To measure larger negative peaks or follow a 
baseline with a strong negative drift, the value needs to be adjusted 
downwards to prevent the signal flat-lining at the bottom of the range. 
However, it should not be changed without good reason, as making it more 
negative will increase baseline noise and reduce the range available for 
measuring positive peaks.
• Advanced - Slit










