Technical data
Agilent 1290 Infinity LC System Manual and Quick Reference 125
Appendix
6
Setting Up a Method using Edit Entire Method
•Stop Time defines the overall time for the separation or run and is sometimes 
referred to as the ‘Run Time’ by some users. This is the time, counted in 
minutes since the injection was made, that the run ends which means that 
data acquisition will stop, the flow, composition and other system settings 
will revert to the initial values for the method and the system will become 
available to make the next injection. It should always be at least as long as 
the last entry in the timetable or else the run will stop and revert to starting 
conditions before the timetable events are complete. Stop Time can be set to 
No Limit in which case the user must manually stop the run. While all the 
system modules have Stop Time parameters, the Stop Time in the pump is 
considered to be the master and the other modules will usually be set to 
follow this value.
•Post Time defines a countdown period after the end of a run during which 
the next injection is prevented. It allows time for the system to 
re-equilibrate after a gradient run. For an isocratic method it can be set to 
Off. For a gradient method the value can be determined experimentally by 
observing the baseline behavior but typically will need time for the delay 
volume of the system plus at least three to five column volumes to be 
flushed through the system.
• Pressure Limits control the behavior of the pump regarding pressure. The 
maximum pressure of the 1290 Infinity pump is 1200 bar but some columns 
can only withstand lower pressure and by setting this value here the 
column will be protected. The pump will generate an error condition if this 
pressure is reached and any run in progress will be stopped and the pump 
switched to standby mode with no flow. Information about the maximum 
pressure for a particular column is supplied with the column. Agilent 
ZORBAX RRHD columns are suitable for operation at 1200 bar. The Low 
Pressure limit is ‘Off’ when the value is zero but at any other value the 
pump will generate an error if the pressure goes below this value during 
operation. This is used as an alternative safeguard when the column is not 
inside a module with a leak sensor or in case the system is pumped dry. A 
value of 10 to 20 bar is typical.










