Technical data
Agilent 1290 Infinity LC System Manual and Quick Reference 123
Appendix
6
Setting Up a Method using Edit Entire Method
Binary Pump (BinPump) tab
Figure 39 Setup Method screen – Binary Pump tab
•Flow sets the flow rate up to 5 ml/min. For the example separation 
0.4 ml/min is used. If the back pressure briefly reaches the maximum 
pressure setting the flow will be reduced for a few seconds to lower the 
pressure but if the pressure continues to be limited in this way an error 
condition will be created and the flow will be stopped.
•Solvents defines the mobile phases that are available and the percentage 
proportions pumped on the two channels, A and B. On each channel a 
drop-down box allows the selection of a solvent from a list so that the pump 
control uses the optimum compressibility settings. This will optimize the 
flow characteristics as described in “How to Configure the Optimum Delay 
Volume” on page 42. A second text box allows a description of the mobile 
phase to be entered. If the solvent selection valve is installed in the pump 
then each channel has two solvent options and the correct option for the 
method is selected with the radio button to the left of the solvent 
description. The pump will form binary mixtures of the selected A and B 