Technical data

Agilent 1290 Infinity LC System Manual and Quick Reference 119
Appendix
6
Setting Up a Method using Edit Entire Method
Autosampler tab (HiP-ALS)
Figure 37 Setup Method screen –High Performance Autosampler tab
•Injection Mode
Injection volume sets the volume to be injected (example 3 µl),
Standard injection indicates that no external needle wash is done,
Injection with needle wash is used to reduce potential carryover. This is the
recommended option and is configured in the next entry.
Needle Wash – if selected above.
Mode determines how the outside of the needle will be rinsed: either
actively in the Flush Port or by dipping into a specified Wash Vial.
Time in seconds the peristaltic pump connected to the flush port will
pump the rinse solvent. It then pumps for a further 15 s to clear the flush
port.