Technical data
Agilent 1290 Infinity LC System Manual and Quick Reference 105
Quick Start Guide
5
Data Analysis
Integrating a Signal
1 Select the Integration Task Tool (see figure below). The Integrate icon and
the Set Integration Events Table icon are highlighted in the figure shown below.
2 Click the Set Integration Events Table icon to open the table as shown.
3 Set Baseline Correction to Advanced for gradient runs.
4 Set Slope Sensitivity to 50. Higher numbers will integrate steeper peaks and
ignore less steep peaks.
5 Set the Peak Width value to the narrowest peak of interest, approximately
0.01 in this case.
6Area Reject and Height Reject can be set to reject the smallest peaks.
7 Click the Integrate icon to update the results using these new settings.