Data Acquisition in Method and Run Control View
Running the Method for a Single Injection
This section shows how to run a single injection of the test mix using the 
conditions entered in the previous section.
Analyses on the ChemStation can be run in two modes:
•Run Method — single injections, for example, in interactive method 
development, using the parameter settings current at the time;
•Run Sequence — automated series of injections from multiple vials, possibly 
with multiple methods. For further details refer to the ChemStation 
manuals.
1 Click the Select Run Method Task icon  .
2 If required method conditions are not currently loaded, select Method > Load 
Method or the icon   under the menu bar to load them.
3 Place the sample vial in position 1 — this is the front position of the 10 x 
2 ml vial positions on the right hand side of the sample tray.
4 Select Run Control > Sample Info, and enter a Subdirectory name (optional), 
Filename, Sample Location (
vial 1), Sample Name and any Comment.
5 If the system is already equilibrated and the baseline is stable, click Run 
Method in the Sample Info page to start the injection. Alternatively, click OK, 
and when ready, click the Start Single Sample button above the system 
diagram.
6 The injection is made, and the chromatogram appears in the Online Plot. The 
data acquisition will stop when the Stop Time is reached.
The chromatogram should look similar to Figure 33 on page 101 — the time 
axis will be longer, however, as this was run under four times faster 
conditions in the next section.
NOTE
If changes have been made to a loaded method and not yet saved, this is indicated by a 
yellow asterisk on the method status icon. The injection can be made without first saving 
those parameters changes. The ChemStation always stores a copy of the acquisition 
parameters in the data file as ACQ.TXT to ensure the preservation of the original method 
parameters.