User`s manual

41
6. Technical Reference
6.1 Calibration
Calibrating the DV2000 involves adjusting a single trimpot only. The pot sets the video output at
the analog monitor output.
Test Equipment
Component digital video pattern generator
75 ohm BNC terminator, 1%
Waveform monitor or ULTECH TV Trigger Mate & 20 MHz oscilloscope
Procedure
qq
qq
q 1. Remove AC line cord. Remove (6) phillips screws from sides of top cover. Remove top
cover.
qq
qq
q 2. Attach AC line cord. Turn unit on. Allow unit and test equipment to warm-up for 15
minutes. Caution: a shock hazard exists with the top cover removed. Do not touch AC
line switch or AC wiring. Apply a stable 1Vpp stairstep pattern to Monitor Video In.
qq
qq
q 3. Adjust R22 trimpot located on the DigMon PC board for 1Vpp at Monitor Video Out
when terminated into 75 ohm.
(end of procedure)
6. Technical Reference
DigMon PC board
R22
ss
ss
s
!
A shock hazard exists with the top cover removed—only qualified
personnel should perform the calibration procedure.
Fig. 6-2. Adjustments