Assembling the camera
The angle of the camera and the adapter on the trinocular head of the
microscope are very important for the panoramic image stitching
function. Please adjust the camera by loosening the screw (the screw
is the one that installs the adapter on the trinocular microscope) (see
the picture on the right, the camera logo is on the front) ).
By horizontally calibrating the slice, adjust the horizontal
movement direction of the camera and the slide to be parallel,
the method is as follows:
1. Place the calibration slice (as shown on the right) horizontally on the stage,
adjust the level of the stage and the level of the slide holder, and observe the
eyepiece field of view to ensure that the calibration slice moves horizontally with
the stage.
2. At this time, observe the camera acquisition page to determine whether the
calibration slice of the acquisition page is also horizontal. If it is not horizontal,
please loosen the screw slightly and turn the connector slightly to adjust the
calibration slice on the camera acquisition page to be horizontal.
Horizontal calibration slice
screw