Owner's manual
S p e c i f i c a t i o n s
Sensor Reconditioning
Over time, the internal sensor may be compromised as a result of pollu-
tants, chemical vapors, and other environmental conditions which can
lead to inaccurate readings. To recondition the internal sensor, please
follow the procedure below:
Dry the logger at 176˚F (80˚C) at <5% RH for 36 hours followed by
70˚ - 90˚F (20˚ - 30˚C) at >74% RH for 48 hours (for re-hydration).
If permanent damage to the internal sensor is suspected, replace the
Logger immediately to ensure accurate readings.
THL1-MAN P. 3
Relative Humidity Overall range 0 to 100%
Accuracy (0 to 20 and 80 to 100%) ±5.0%
Accuracy (20 to 40 and 60 to 80%) ±3.5%
Accuracy (40 to 60%) 3.0%
Temperature Overall range -40˚ to 158˚F
(-40˚ to 70˚C)
Accuracy (-40˚ to +14 and 104˚ to 158˚F) ±3.6˚F
Accuracy (+14˚ to +104˚F) ±1.8˚F
Accuracy (-40˚ to -10˚ and +40˚ to +70˚C) ±2˚C
Accuracy (-10˚ to +40˚C) ±1˚C
Dew Point Overall range -40˚ to 158˚F
Temperature (-40˚ to 70˚C)
Accuracy (25˚C, 40 to 100% RH) ±4.0˚F (2.0˚C)
Logging Rate Selectable sampling interval: From 2 seconds up to
24 hours.
Operating -31˚ to 175˚F (-35˚ to 80˚C)
Temperature
Battery Type 306V lithium (1/2AA) (SAFT LS14250, Tadiran TL-5101
or equivalent)
Battery Life 1 year (typ.) depending on logging rate, ambient
temperature and use of alarm LEDs.
Dimensions 101 x 25 x 23 mm (4 x 1 x 9”)
Weight 172 g (6 oz)