User Manual

aNalyzer coNNectioNS
NOTE: Take care when inserting the temperature
probes as the pins are polarized. Insert with the smaller pin into input
markedpositive.(+)
WARNING!
Turning the pump off while the probe is in the flue
will leave toxic gases inside the analyzer. Once data
has been read, it is advisable to purge the
unit with fresh air as soon as possible. To do this,
with the probe removed from the flue, turn ON the
pump. Always allow the readings to return to zero
(0.0forCO2)priortoshuttingtheunitoff.
WARNING!
Theprobewillbehotfromfluegases.Removetheprobefromtheflueandallowittocoolnaturally.Do
not immerse the probe in water, as this will be drawn into the analyzer and damage the pump and sensors. Once the
probe is removed from the flue and the readings have returned to ambient levels rotate the selector
to“OFF”.Theinstrumentwillcountdownfrom30toswitchoff.
fiNiShiNG up
• Remove the probe from the flue and allow the analyzer to purge with fresh air until readings return to zero.
-CO2to0.0%,COtoZero(Be careful as the probe tip will be HOT)
• Drainwatertrap
 • Checkparticlefilter
• Rotateselectorto“OFF”
NOTE:Toprotectthesensor“HIGH”willbedisplayedandthezerocountdownwillrepeatuntil
COmeasuredislessthan20ppm.
SmartBell pluS overview
OFF
L
O
i
l
P
r
o
p
a
n
e
N
a
t
G
a
s
H
O
i
l
W
o
o
d
S
t
a
n
d
b
y
C
O
C
O
2
O
2
X
-
A
i
r
T
F
l
u
e
E
f
f
˚
F
/
˚
C
˚C ˚F
0000
LCD Display
Selector
Dial
Water
Trap
Fuel/Test Type
Indicator
2