User Manual
Safety NoteS
Before using this meter, read all safety information carefully. In this manual the word “WARNING” is used to indicate
conditions or actions that may pose physical hazards to the user. The word “CAUTION” is used to indicate conditions or
actions that may damage this instrument.
WARNING!
This analyzer extracts combustion gases that may be toxic in relatively low concentrations. These gases are exhausted
from the back of the instrument. This instrument must only be used in well-ventilated locations.It must only be used by
trained and competent persons after due consideration of all the potential hazards.
GettiNG Started
• Cleanparticlefilter
• Watertrapandprobelineareemptyofwater
• Poweronandzero
• Allhoseandthermocoupleconnectionsareproperlysecured
• FluegasprobeissamplingambientFRESHair
• Watertrapisfittedcorrectlytotheinstrument
• Fluetemperatureplugisconnected-Onlyifsettinginlettemperature
SettiNG temperature Scale
Tochangethetemperaturescalefrom°Fto°C(orback)performthefollowingsteps;
•StartwiththeSmartBellPLUSOff
•Rotatetheselectorto“˚F/˚C”
o The current scale will initially be displayed
o The unit will switch to the alternative scale
•Rotatetheselectortothedesiredfuelforyourtesting
•Allowtheunittocompletethezeroprocess
Note: These steps are only required one time, or when the scale needs to be changed.
SettiNG iNlet temperature
To correctly calculate net temperature and combustion efficiency it is important to have the correct inlet tempera-
ture set. Two methods are available to properly establish the inlet temperature
• Non-ductedsystems.IftheSmartBellPLUSisbeingusedonasystemthatusescombustionairfromthespace
neartheappliancesimplyturntheanalyzeronwithouttheflueprobeconnected.TheSmartBellPLUSwilluseanambi-
ent sensor inside the handset as the inlet temperature to determine net temperature.
• Ductedsystems.IftheSmartBellPLUSisbeingusedonasystemthatbringsinoutsideairforcombustionairit
is important to set the inlet temperature prior to taking flue gas readings. To properly set inlet perform the following steps
during startup.
1.ConnectthefluetemperatureconnectoronlytotheSmartBellPLUS
2.Placetheflueprobeinthecombustionairinletstream(youwillneedtodrilla1/4”holeforaccess)
3.TurnontheSmartBellPLUSandallowtocompletethezeroprocess.
4.ConnectthefluegasconnectortotheSmartBellandproceedtocombustionmeasurements
•Thetemperaturemeasuredduringthezerocountdownhasnowbeenstored,andwillbeusedtodetermine
the net temperature.
SelectiNG fuel
WhenpoweringtheSmartBellPLUSon,simplyrotatetheselectortothedesiredfuel.Thedisplaywillshowthisdisplay,
and then it will complete the zero process.
Note:Ifyouwishtousethesamefuelaspreviouslyselectedrotatetheselectorto“StandBy”atpoweron.Thefuelinuse
will be displayed and then blink. If the fuel is not correct rotate the selector to the proper choice before the Zero
countdown starts.
OWNERS MANUAL & MAINTENANCE