User Guide
8
AC/DC Voltage: <1000V AC/DC
AC VOLTS
Ranges
Accuracy
Resolution
Overload
Protection
45Hz to 500Hz 500Hz to 5kHz 5kHz to 20kHz
6.000V
±(0.75% +5 dgts) ±(2.0% +8 dgts)
±(2.0% +20 dgts)
0.001V
1000V
60.00V 0.01V
600.0V
Unspecified
0.1V
1000V ±(0.75% +8 dgts) ±(2.0% +8 dgts)* 1V
Minimum sensitivity: 0.5V AC (auto selection mode), *Accuracy for 500Hz to1kHz only
DC VOLTS
Ranges Accuracy Resolution Overload Protection
6.000V to 1000V ±(0.2% +5 dgts) 0.001V to 1V 1000V
Minimum sensitivity: 0.5V DC (auto selection mode)
AC MILLIVOLTS
Ranges
Accuracy
Resolution
Overload
Protection
45Hz to 500Hz 500Hz to 5kHz 5kHz to 20kHz
600.0mV ±(0.75% +5 dgts) ±(2.0% +8 dgts) ±(2.0% +20 dgts) 0.1mV 600V
Auto selection mode is not available.
DC MILLIVOLTS
Ranges Accuracy Resolution Overload Protection
600.0mV ±(0.2% +5 dgts) 0.1mV 600V
Auto selection mode is not available.
Note: Meter automatically selects AC or DC
WARNING
• Use CATIV rated Test leads or higher.
• Do not attempt to measure more than 1000V AC/DC.
• Do not exceed 25 volts AC or 60 volts DC – RMS at either the common
or multifunction input ports as measured from earth ground.
Features:
RANGE
MIN/MAX
HOLD
SEL
RANGE
SEL
Temp
A
True RMS
MIN/MAX
HOLD
FAST
Low Z
V
Low Z
Apo
µA
mAµA
A
mA
10A
MAX
FUSED
400mA
MAX
FUSED
MAX
CAT IV
600V
CAT III
1000V
DM525
mV
°F°C
REL
Hz
Hz %
W
PF
0 10 20 30 40 50 60
LINK
Low Z
µA
A
mA
mV
°F°C
Hz %
W
PF
V
Apo
AT
DM525
0 10 20 30 40 50 60