DM525 True RMS 1000V Digital Multimeter w/Temperature INSTRUCTION MANUAL ENGLISH DM525 Low Z Apo V 0 10 20 RANGE 30 40 MIN/MAX 50 60 HOLD FAST SEL REL °F°C Hz % mV LINK W PF Low Z µA A mA Hz Temp True RMS A mAµA 10A MAX FUSED 400mA MAX FUSED 1-800-547-5740 www.ueitest.com • email: info@ueitest.
TABLE OF CONTENTS FUNCTIONS .................................................................................................................................................. 4 FEATURES ..................................................................................................................................................... 4 GENERAL SPECIFICATIONS ..................................................................................................................... 4 IMPORTANT SAFETY WARNINGS .....
FUNCTIONS • 1000V AC/DC • Frequency 999.9 kHz • Resistance 50MΩ • Microamps • Diode test • Milliamps • Audible continuity • Duty cycle 99.
IMPORTANT SAFETY WARNINGS WARNING Read entire Safety Notes section regarding potential hazard and proper instructions before using this meter. In this manual the word “WARNING” is used to indicate conditions or actions that may pose physical hazards to the user. The word “CAUTION” is used to indicate conditions or actions that may damage this instrument. WARNING To ensure safe operation and service of the tester, follow these instructions.
OVERVIEW A L DM525 B C D E F G H Low Z Apo V 0 10 20 RANGE 30 40 MIN/MAX 50 60 HOLD FAST SEL REL °F°C Hz % mV M I J N O LINK W PF Low Z P µA A mA Q Hz Temp True RMS A mAµA 10A MAX FUSED 400mA MAX FUSED MAX CAT IV 600V CAT III 1000V K A. B. C. D. Worklight Apo: Auto power off after 30 minutes of use. Press and hold the HOLD button while turning the meter on to disable Apo.
OVERVIEW (CONT.) K. Category Max Indicator: Maximum CAT Rating for fused input jacks. • Multifunction input port used for measuring: AC or DC volts, AC or DC mV, Temperature, Hertz, Duty Cycle, Watts, Resistance, Continuity, Diode, Capacitance, AC or DC Amps, AC or DC µA, AC or DC mA. • Use CATIV test leads or higher L. Recess for magnetic hanger M. Test Lead Holders N. Protective Rubber Boot O. Battery Cover (under protective rubber boot) P. Kick Stand Q.
AC/DC Voltage: <1000V AC/DC Note: Meter automatically selects AC or DC DM525 Low Z WARNING • Use CATIV rated Test leads or higher. • Do not attempt to measure more than 1000V AC/DC. • Do not exceed 25 volts AC or 60 volts DC – RMS at either the common or multifunction input ports as measured from earth ground.
Resistance: <50MΩ DM525 Low Z SEL Apo V AT 0 10 20 RANGE 30 40 50 MIN/MAX 60 • • • • HOLD FAST SEL REL °F°C LINK Hz % mV Hz W PF Low ZZ Low µA A mA Default = Press x1 = Press x2 = Press x3 = Hz Temp Features: True RMS A mAµA MAX CAT IV 600V CAT III 1000V 400mA MAX FUSED 10A MAX FUSED RANGE MIN/MAX HOLD WARNING • Do not measure resistance on a live circuit. Ranges 600.0Ω 6.000kΩ 60.00kΩ 600.0kΩ 6.000MΩ 50.00MΩ Accuracy Resolution 0.1Ω 0.001kΩ 0.01kΩ 0.1kΩ 0.001MΩ 0.
Diode GOOD DIODE BAD DIODE DM525 Forward Bias Displays approx. voltage drop Open Diode Displays "OL" Both directions SEL Low Z Apo V 0 10 20 RANGE 30 40 MIN/MAX 50 60 HOLD FAST SEL REL °F°C Hz % mV Hz Reverse Bias Displays "OL" LINK W PF Low ZZ Low µA • • • • 400mA MAX FUSED V '0' Both directions (shorted) Hz Temp 10A MAX FUSED Apo • Forward voltage drop if forward biased. • “OL” if reverse biased.
AC/DC Microamps: <6000µA AC/DC Milliamps: 400mA DM525 DM525 DM525 Low Z Apo µ AT 00 10 10 20 20 RANGE 30 30 40 40 MIN/MAX Low Z Apo A V m 50 50 60 60 0 HOLD 10 20 30 RANGE 40 MIN/MAX 50 FAST SEL °F°C °F°C mV Hz Hz % % mV LINK REL W W PF PF Low Z °F°C Hz % mV Hz µA A A mA mA A mA Hz Temp Hz Temp True RMS mAµA 10A MAX FUSED 400mA MAX FUSED LINK W PF Low ZZ Low µA µA A 60 HOLD FAST SEL REL VA AT True RMS A MAX CAT IV 600V CAT III 1000V µA mAµA 10A MAX FUS
AC/DC Amps: <10A DM525 Low Z Apo VA AT 0 10 20 RANGE 30 40 MIN/MAX 50 60 HOLD FAST SEL REL °F°C Hz % mV Low Z LINK W PF µA A mA Hz Temp WARNING Keep hands below guard when measuring current levels. • Do not attempt to measure more than 10A AC. Features: RANGE MIN/MAX HOLD True RMS A mAµA 10A MAX FUSED 400mA MAX FUSED MAX CAT IV 600V CAT III 1000V A SEL AC A Ranges Accuracy 45Hz to 500Hz 500Hz to 5kHz 6.000A 10.00A ±(1.2% +5 dgts) ±(2.
Temperature C˚/F˚ • • • • Default = mV Press x1 = mV Press x2 = ˚F Press x3 = ˚C DM525 DM525 Low Z SEL Apo Apo V AT 00 10 10 20 20 RANGE 30 30 40 40 MIN/MAX 50 50 6060 HOLD FAST SEL REL °F°C Hz % mV Hz LINK W PF Low ZZ Low µA A mA • Press Select button to change scale between Fahrenheit and Celsius. Hz Temp True RMS A Features: MIN/MAX HOLD mAµA 10A MAX FUSED SEL MAX CAT IV 600V CAT III 1000V 400mA MAX FUSED °F Ranges -328˚ to 999˚F 1000˚ to 2462˚F Accuracy ±(1.5% + 3.
Frequency/Duty Cycle DM525 Low Z Apo SEL V AT Hz 0 • Default = Hz • Press x1 = % 10 20 RANGE 30 40 MIN/MAX 50 60 HOLD FAST SEL REL °F°C Hz % mV Hz LINK W PF Low ZZ Low µA A mA Hz Temp True RMS A mAµA 10A MAX FUSED Features: RANGE MIN/MAX HOLD 400mA MAX FUSED MAX CAT IV 600V CAT III 1000V V SEL Frequency Ranges Accuracy Resolution Overload Protection 99.99Hz to 999.9kHz ±(0.05% +3 dgts) 0.01Hz to 0.1kHz 600V Duty Cycle Ranges Accuracy Overload Protection 1.
Watt (Power Factor) Measurement of Active, Apparent and Reactive Power; W,VA, VAr The current path be measured directly (up to 10A maximum; up to 16A briefly for a maximum of 30 seconds) or with the help of current transformers or current clamp transformers. The meter automatically selects the range which allows for the highest possible resolution of the applied quantities.
4mm Test Lead Notes CATV 600V Measurement Locations 4mm • Ensure the test lead shield is pressed firmly in place. Failure to use the CATIV shield increases arc-flash risk. CATII 1000V Measurement Locations 18mm 18mm • CAT IV shields may be removed for CAT II locations. This will allow testing on recessed conductors such as standard wall outlets. Take care not to lose the shields.
CONNECTING AND USING THE APP • • • • • Search for App as, “525 DMM”. Compatible with iPhone 4X and up running iOS7 or higher, Galaxy S4, Nexus5, HTC One running Android 4.4 or higher. To install ot search on iPad use “iPhone only” to find App. Press and hold “LINK” button on the meter to activate wireless “BT”. Open App. Meter will connect automatically. Menu • Press “ ” to connect, disconnect, and access settings. Settings • General settings adjust button sound, vibrate and refresh rate.
Logs • Press “ ” to view recorded data. • • Press the entry you wish to view (yyyy-mm-dd hh:mm:ss). Functions are noted underneath respectively AMP-AMP (TOP-BOTTOM) Display. • Press “ “ button for summary. • Press “ “ button for sample data. • Press “ “ button to export data via email in .csv, .png or .jpg format.
Graph • Press “ ” to view trending data in real time during measurement. FCC/IC INFORMATION NOTE: This device complies with Part 15 of the FCC Rules and CAN ICES-3(A). Operation is subject to the following two conditions: (1) this device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operations.
WARRANTY The DM525 is warranted to be free from defects in materials and workmanship for a period of 2 years from the date of purchase. If within the warranty period your instrument should become inoperative from such defects, the unit will be repaired or replaced at UEi’s option. This warranty covers normal use and does not cover damage which occurs in shipment or failure which results from alteration, tampering, accident, misuse, abuse, neglect or improper maintenance.