DL489 Commemorative Edition True RMS Digital Clamp-On Meter INSTRUCTION MANUAL ENGLISH CAT III 600A 600V μA z/ D MF D VH 600V CAT III MAX/ MIN PEAK Tem A F°C p/° NCV u ty RANGE DC Zero INRUSH DL489 True RMS 1000V CAT II T1 30V T2 CATII 1000V CATIII 600V TEMP MAX 1-800-547-5740 www.ueitest.com • email: info@ueitest.
TABLE OF CONTENTS FEATURES ..............................................................................................3 GENERAL SPECIFICATIONS ..............................................................3 IMPORTANT SAFETY WARNINGS ............................................. 3 - 4 CATEGORY DEFINITIONS....................................................................4 SYMBOLS ..............................................................................................5 OVERVIEW .....................
FEATURES • True RMS • Min/Max • 600A AC/DC • Peak hold • 750V AC/1000V DC • Visible high-voltage alert • Resistance 60MΩ • Manual ranging option • Capacitance 2000µF • Worklight/ back lit display • AC/DC microamps 2000µA • Low battery indicator • Frequency 99.
IMPORTANT SAFETY WARNINGS (CONT.) • Use caution when working with voltages above 60V DC or 25V AC RMS. Such voltages pose shock hazards. • To avoid false readings that can lead to electrical shock, replace batteries if a low battery indicator appears. • Unless measuring voltage or current, shut off and lockout power before measuring resistance or capacitance. • Always adhere to national and local safety codes.
SYMBOLS AC (Alternating current) DC (Direct current) Negative DC AC/DC Voltage or Current Auto-ranging Overload: Range Exceeded Auto power off Active Non-Contact Voltage Low Battery Hold/Capture Value Minimum measured value displayed Maximum measured value displayed Duty Cycle Hertz/Frequency Voltage Amperage Ohms/Resistance Diode Capacitance mode in nanofarads or microfarads µF Microfarad Degrees Fahrenheit Degrees Celsius Continuity Mega (x106 or 1,000,000) Milli (x10-3 or 0.
OVERVIEW (CONT.) A. Clamp: Measure inductive AC/DC current. Opens to 1.25” (31.7mm). B. Conductor Alignment Marks: Use to aid the visual alignment of a conductor when measuring inductive amperage. Greatest accuracy is achieved when the conductor inside the clamp is centered at the intersection of these marks. C. Wire Separation Tab/ NCV sensor: Use to isolate an individual wire from a bundle for testing. NCV sensor detects live voltage. D. Test Lead Holder: Used for hands-free use of the test probes. E.
S. DC Zero/ Inrush Button: • Press to Zero the DC clamp reading. • Press and hold to enter LRA Inrush mode (must be in AC amp mode first). • Press and hold again to return to live readings. T. Display: • High contrast dual backlit display. • Amps (AC/DC) reading will always display on upper display. U. K-Type Temperature Probe Inputs: T1 (Upper) and T2 (Lower) V: Input Jack Lock: Switch to use Temperature or Test lead inputs W. Category Max Indicator: Maximum CAT Rating for input jacks.
DC Amps: <600A – Jaw CAT III 600A 600V Tem μA A MFD VH z/ D F°C p/° NCV MAX/ MIN PEAK Single Conductor Only u ty RANGE DC Zero INRUSH • • • • DL489 True RMS T1 30V T2 CATII 1000V CATIII 600V Default = (upper) Press x1 = µA (lower) Press x2 = µA (lower) Press and hold = (upper) TEMP MAX • Do not attempt to measure more than 600A DC. Features: MAX/ MIN PEAK RANGE HOLD Range Resolution Accuracy Overload Protection 60A 600A 0.01A 0.1A ±(2.
Zero DC Amps CAT III 600A 600V Tem A RANGE DC Zero INRUSH MFD VH z/ D F°C p/° μA NCV MAX/ MIN PEAK u ty DL489 True RMS T1 • Default = DC Zero • Press and Hold = LRA Inrush 30V T2 TEMP CATII 1000V CATIII 600V MAX Select DC current. • Press to zero any offset in Amps DC. • Used to monitor change from present displayed value. • Required during DC Amps measurement to establish a zero level. WARNING Do not use DC Zero mode at amps greater than 600A DC or 2000µA DC.
Temperature F˚/C˚ CAT III 600A 600V Tem μA MFD VH u ty F°C p/° Tem A z/ D F°C p/° NCV MAX/ MIN PEAK RANGE DC Zero INRUSH • • • • DL489 True RMS TEMP T1 30V T2 CATII 1000V CATIII 600V TEMP Default = T1 Press x1 = T2 Press x2 = T1-T2 Press and hold = MAX Range Resolution -328˚F to 999˚F (-200˚ to 999˚C) 0.1°F (0.1˚C) 1000˚F to 2462˚F 1°F (1˚C) (1000˚ to 1350˚C) Accuracy Overload Protection ±(1.0% +3.6˚F) ±(1.0% + 2.
Resistance: < 60MΩ CAT III 600A 600V Tem μA A MFD VH z/ D F°C p/° NCV MAX/ MIN PEAK u ty RANGE • • • • DC Zero INRUSH DL489 True RMS T1 30V T2 CATII 1000V CATIII 600V Features: Default = Press x1 = Press x2 = Press x3 = TEMP MAX RANGE HOLD Range 600Ω 6kΩ 60kΩ 600kΩ 6MΩ 60MΩ MAX/ MIN PEAK Resolution 0.1Ω 1Ω 10Ω 100Ω 1kΩ 0.01MΩ Accuracy Overload Protection ± (0.8% +3 dgts) 600V RMS ± (1.
Diode GOOD DIODE • • • • CAT III 600A 600V Reverse Bias Displays "OL" Tem A MFD VH z/ D RANGE F°C p/° μA NCV MAX/ MIN PEAK u ty Default = Press x1 = Press x2 = Press x3 = DC Zero INRUSH DL489 True RMS Forward Bias Displays approx. voltage drop T1 30V T2 CATII 1000V CATIII 600V TEMP MAX BAD DIODE or Open Diode Displays "OL" Both directions • Forward voltage drop if forward biased. '0' Both directions • “O.L.” if reverse biased.
Voltage: AC 750V / DC 1000V CAT III 600A 600V Tem μA A VH z/ D MFD NCV MAX/ MIN PEAK F°C p/° High Voltage Alert u ty RANGE DC Zero INRUSH DL489 True RMS T1 • • • • Default = Press x1 = Press and hold = Press and hold x1 = 30V T2 CATII 1000V CATIII 600V TEMP MAX WARNING Use CATIII rated test leads or higher. Do not attempt to measure more than 750V AC/1000V DC. Keep hands below line when measuring high current levels.
Frequency (Hz) / Duty Cycle CAT III 600A 600V Tem μA A MFD VH z/ D F°C p/° NCV MAX/ MIN PEAK u ty RANGE • • • • DC Zero INRUSH DL489 True RMS T1 Default = Press x1 = Press and hold = Press and hold x1 = 30V T2 CATII 1000V CATIII 600V TEMP MAX Use CAT III rated leads or higher. Press the AC/DC volts button to select AC voltage, press and hold the button for Frequency and Duty Cycle modes. WARNING Do not attempt to measure more than 750V AC/1000V DC.
LRA Inrush CAT III 600A 600V Tem μA A VH z/ D MFD F°C p/° NCV MAX/ MIN PEAK u ty RANGE DC Zero INRUSH DL489 True RMS • Default = DC Zero • Press and hold = LRA Inrush T1 30V T2 CATII 1000V CATIII 600V TEMP MAX The UEi LRA Inrush is programmed to properly capture the starting current for compressor motors. • Select AC Amps. • Select the range capable of capturing the maximum value. • Press and hold the DC Zero/ LRA Inrush button – INRUSH will now be shown on the screen.
WARRANTY The DL489 is warranted to be free from defects in materials and workmanship for a period of 2 years from the date of purchase. If within the warranty period your instrument should become inoperative from such defects, the unit will be repaired or replaced at UEi’s option. This warranty covers normal use and does not cover damage which occurs in shipment or failure which results from alteration, tampering, accident, misuse, abuse, neglect or improper maintenance.