Install Instructions

DL479
NCV
°C°F
CAT IV 600V
CAT III 1000V
MAX
µA
A
CAT Ill
1000V 600A
True RMS
Hz / Duty
MAX/MIN
AC/DC Voltage: <750V AC/600V DC
Default =
Press x1 =
WARNING
Use CAT III rated Test leads or higher.
Do not attempt to measure more than 750V AC/600V DC.
Keep hands below line when measuring high current levels.
Do not exceed 600 volts AC or DC – RMS at either the common or
multifunction input ports as measured from earth ground.
Select AC or DC Voltage.
WARNING
High Voltage indicator will display and audible alert will sound over 600V
AC/DC and High Voltage indicator will display (without audible alert) over
30V AC/DC.
Features:
MAX/MIN
HOLD
DC Volts
Range Resolution Accuracy Overload Protection
600mV 0.1mV
±(0.5% +4 dgts) 1000V RMS
6V 1mV
60V 10mV
600V 100mV
AC Volts (45Hz to 400Hz)
Range Resolution Accuracy Overload Protection
600mV 0.1mV
±(1.0% + 3 dgts) 1000V RMS
6V 1mV
60V 10mV
600V 100mV
750V 1.0V
True RMS Frequency Range: Square 50Hz to 170Hz
Sine 50Hz to 400Hz
Bandwidth: Sine = 0.5% error at 1.5kHz (max)
Square = 0.5% error at 0.1kHz (max)
Triangle = 0.5% error at 1.2kHz (max)
DL479
NCV
°C°F
CAT IV 600V
CAT III 1000V
MAX
µA
A
CAT Ill
1000V 600A
True RMS
Hz / Duty
MAX/MIN
Range Resolution Accuracy Overload Protection
600Ω 0.1Ω
±(0.8% +3 dgts)
600V RMS
6kΩ
60kΩ 10Ω
600kΩ 100Ω
6MΩ 0.001MΩ
60MΩ 0.01MΩ ±(1.2% +3 dgts)
Resistance: <60MΩ
Features:
MAX/MIN
HOLD
WARNING
Do not measure resistance on a live circuit.
Default =
Press x1 =
Press x2 =
Press x3 =
Buzzer sounds at less than 40Ω.
WARNING
Do not measure resistance on a live circuit.
Features:
HOLD
Open Circuit V Response Time Overload Protection
<1.0V <50ms 600V RMS
Default =
Press x1 =
Press x2 =
Press x3 =
Continuity
DL479
NCV
°C°F
CAT IV 600V
CAT III 1000V
MAX
µA
A
CAT Ill
1000V 600A
True RMS
Hz / Duty
MAX/MIN
DL479
NCV
°C°F
CAT IV 600V
CAT III 100 0V
MAX
µA
A
CAT Ill
1000V 600A
True RMS
Hz / Duty
MAX/MIN
Reverse Bias
Displays "OL"
Forward Bias
Displays approx.
voltage drop
Default =
Press x1 =
Press x2 =
Press x3 =
Diode
GOOD DIODE
Forward voltage drop if forward biased.
“O.L.” if reverse biased.
Features:
HOLD
BAD DIODE
'0' Both directions
(Shorted)
or
Range Open Circuit V Test Current Overload Protection
4.0V <3.0V DC 1.30mA 600V RMS
Default =
Press x1 =
Press x2 =
Press x3 =
Capacitance (MFD)
DL479
NCV
°C°F
CAT IV 600V
CAT III 100 0V
MAX
µA
A
CAT Ill
1000V 600A
True RMS
Hz / Duty
MAX/MIN
Features:
MAX/MIN
HOLD
Range Resolution Accuracy Overload Protection
60nF 0.01nF
±(3.0% +5 dgts) 600V RMS
600nF 0.1nF
6.000µF 0.001µF
60.00µF 0.01µF
600.0µF 0.1µF
2000µF 1µF
Single
Conductor
Only
DL469
NCV
CAT III 100 0V
MAX
µA
A
CAT Ill
1000V 400A
True RMS
AC Amps: < 600A
DL479
NCV
°C°F
CAT IV 600V
CAT III 100 0V
MAX
µA
A
CAT Ill
1000V 600A
True RMS
Hz / Duty
MAX/MIN
DL469
NCV
CAT III 100 0V
MAX
µA
A
CAT Ill
1000V 400A
True RMS
Center wire in conductor alignment marks for best accuracy.
Opposing currents cancel (use line-splitter when necessary).
WARNING
Keep hands below guard when measuring high current levels.
Do not attempt to measure more than 600A AC.
Features:
MAX/MIN
HOLD
AC Amps Measurement – Jaw input
Range Resolution Accuracy Overload Protection
60A 0.01A
±(2.0% +8 dgts) 600V RMS
600A 0.1A
True RMS Frequency Range: Sine 50Hz to 400Hz
Frequency width: 60Hz to 400Hz: 5% to 95%
Frequency width: 400Hz to 4kHz: 15% to 85%
WARNING
Do not attempt to measure more than 2000µA DC
Features:
MAX/MIN
HOLD
DC Low Amps – Test Lead input
Range Resolution Accuracy Overload Protection
600µA 0.1µA
±(1.2% +3 dgts) 2000µA/600V RMS
2000µA 1µA
True RMS: 45Hz to 400Hz
DC Low Amps: < 2000µA
DL479
NCV
°C°F
CAT IV 600V
CAT III 1000V
MAX
µA
A
CAT Ill
1000V 600A
True RMS
Hz / Duty
MAX/MIN
M
Temperature C˚/F˚
DL479
NCV
°C°F
CAT IV 600V
CAT III 1000V
MAX
µA
A
CAT Ill
1000V 600A
True RMS
Hz / Duty
MAX/MIN
Disconnect test lead probes from voltage source and meter.
Press Select button to change between Fahrenheit and Celsius
Features:
MAX/MIN
HOLD
Range Resolution Accuracy Overload Protection
14° to 752°F 0.1°F ±(1.5% +3.6°F )
600V RMS
-10° to 400°C 0.1°C ±(1.5% +2 °C)
Sensor must be thermocouple type.
Stated accuracy does not account for themocople accuaracy.
Sensitivity >2Vpp RMS
Default =
Press x1 =
v
com
K
TYPE
°F
A
µ
A
Open Diode
Displays "OL"
Both directions
Non-Contact Voltage
DL479
NCV
°C°F
CAT IV 600V
CAT III 1000V
MAX
µA
A
CAT Ill
1000V 600A
True RMS
Hz / Duty
MAX/MIN
Select NCV mode and move the tip of the clamp meter near voltage source.
Both an Audible and Visual alert will indicate voltage.
Non-Contact Voltage Detection is used to detect power with sensor located
in the tip of the clamp head indicates positive response with both an Audible
and Visual alert.
Do not use non-contact voltage detector to determine if there is current on
the wire. Detection operation could be affected by socket design, insulation
thickness, type or other factors.
Voltage indicator light may also light when voltage is present on the meter’s
input jack or from an external interference such as motors, flashlights, etc.
NCV Sensor
located in tip
On Voltage
Approx. 25V AC
Frequency (Hz) / Duty Cycle
DL479
NCV
°C°F
CAT IV 600V
CAT III 1000V
MAX
µA
A
CAT Ill
1000V 600A
True RMS
Hz / Duty
MAX/MIN
• Use CAT III test leads or higher.
Set Function dial to AC/DC Amps, select AC Amps, press and hold the
Select button for Frequency and Duty Cycle modes.
• WARNING
• Do not attempt to measure more than 750V AC/600V DC.
Features:
HOLD
Frequency
Range Resolution Accuracy Overload Protection
99.99Hz 0.01Hz
±(0.1% +3 dgts) 600V RMS999.9Hz 0.1Hz
9.999kHz 0.001kHz
Duty Cycle
Range Resolution Accuracy Overload Protection
0.5% to 95%
60Hz to 400Hz
0.1%
± (0.2% per kHz
+2.0% +2 dgts)
600V RMS
15% to 85%
400Hz to 2kHz
Frequency Width: 60Hz to 400Hz: 0.5% to 95%
400Hz to 2kHz: 15% to 85%
Sensitivity: >6Vpp RMS
• Default =
• Press and hold =
• Press =
Hz