User Guide

Auto On/APO
Press any button to power on meter.
Press and hold AC/DC Volts/Power button to power off meter.
Auto power off (Apo) after 10 minutes of inactivity.
Press and hold the Hold button when meter is in Hold mode to disable Auto
power off mode. Press and hold again to enable Auto power off mode.
Default =
Press x1 =
Press x2 =
Default =
Press x1 =
Press x2 =
CAT III
600V
V
Ω
DL200
600V CAT III 200A
HOLD
HOLD
AC/DC Voltage: <600V AC/DC
WARNING
Do not attempt to measure more than 600V AC/DC.
Keep hands below line when measuring high current levels.
Do not exceed 600 volts AC or DC – RMS on either the common or
multifunction test leads as measured from earth ground.
Select AC or DC Voltage.
Features:
HOLD
AC Volts
Ranges Resolution Accuracy Overload Protection
1V to 199.9V 0.1V
±(1.5% + 5 dgts) 600V
200V to 600V 1V
DC Volts
Ranges Resolution Accuracy Overload Protection
1V to 19.9V 0.01V
±(1.5% + 3 dgts) 600V20V to 199.9V 0.1V
200V to 600V 1V
Minimum measurement AC mV: 500mV/DC mV:40mV
Response time (input; AC 100V): 1 sec
Max displayed value: 660V AC/DC
High warning symbol input voltage > AC30V/ > DC30V
Meter beeps input voltage > AC 601V/ > DC 601V
Input impedance: 10MΩ
Frequency Range: sine wave: 45Hz to 400Hz
Overload protection: 600V AC RMS or 600V DC
Resistance: <2000Ω
Features:
HOLD
WARNING
Do not measure resistance on a live circuit.
Resistance
Ranges Resolution Accuracy Overload Protection
200Ω to 2000Ω 0.1Ω to 1Ω ±(0.8% + 4 dgts) 600V
Max displayed value: 2200Ω
Overload protection: 600V AC RMS or 600V DC
Default =
Press x1 =
Press x2 =
CAT III
600V
V
Ω
DL200
600V CAT III 200A
HOLD
CAT III
600V
V
Ω
DL200
600V CAT III 200A
V