  Displays estimated battery life.
If battery voltage falls below a pre-set limit, the display flashes .” every 10 seconds.


Check for the following:
The particle filter is not dirty
The water trap and probe line are empty of water
All hose and thermocouple connections are properly made
The probe is sampling ambient FRESH air
The water trap and drain plug are fitted correctly to the instrument
The temperature probe is connected if required

Place the CO95’s probe into the appliance’s sampling point. The ideal sampling point is at least two flue
diameters downstream of any bend.
Put the probe tip in the flue center. With balanced flue, make sure the probe is positioned far enough into the
flue so no air can “back flush” into the probe.
Make sure you do not exceed the meter’s operating specifications.
In particular:
Do not exceed the probe’s maximum temperature
Do not exceed the meter’s internal temperature operating range
Do not put the meter on a hot surface
Do not exceed the water trap’s levels
Do not let the meter’s particle filter become dirty and blocked
Press “” first to “” the readings before printing, then press “”.