User guide

Probe Configuration
Operating Instructions
Before Use Each Time
The particle filter is not dirty
The water trap and probe line are empty of water
All hose and thermocouple connections are properly made
The flue gas probe is sampling ambient FRESH air
The water trap is fitted correctly to the instrument
The flue temperature plug is connected
The inlet temperature probe is connected if required
Switch ON the instrument by pressing “ “.
After switch-on, the analyzer will scroll through the following
information while performing a zero countdown:
Currently Set Date
Currently Set Time
Remaining Battery Level
Fuel Selected
Model and Analyzer Firmware Level
NOTE: The count begins at 59 seconds, and will display the
parameter selected with the rotary knob when the sensors are
detected as stable. If the analyzer will not auto-zero, the sensors are
in need of replacement and the unit should be returned to the
authorized service center.
C O / C O 2 : The CO / CO2 ratio is the ratio of measured CO divided
by calculated CO 2 .
It gives an indication of:
H o w good a gas sample the instrument is reading
H o w clean the boiler is running
For example: A new or clean domestic boiler will display a ratio of less
than 0.004, a unit in need of cleaning 0.004 - 0.008 and a unit in need
of major overhaul will show greater than 0.008.
This only shows a reading when a combustion test is being carried out.
- - - - is displayed while in fresh air.
: Displays the approximate battery level as follow s :
Full battery level
B a t t e ry a t 75 %
B a t t e ry a t 50 %
B a t t e ry a t 25 %
When the display flashes this, it indicates the batteries are at less than
10% of charge and should be replaced, readings may be affected if the
analyzer is used with low power batteries.
DATE : Date shown as day, month and year. The order can be
changed using the menu function. Date is stored with a
combustion test.
TIME : The time is shown in hours and minutes, these details are
stored with each combustion test.
NOT E : When changing the batteries on the instrument the
memory will store the date and time for up to one minute,
if outside this time it may be necessary to re-enter the details.
C50/C75-MAN P. 2