C165+ Quick Start Guide BACKED BY O2/Eff CO/CO2 Temp Exch Test Aux Status ROOM CO MP PRS v3 TE LINK CO 2 CO C165+ Com busti o n A n al r yze 1-800-547-5740 • Fax: (503) 643-6322 www.ueitest.com • Email: info@ueitest.
C165+ OVERVIEW A. B. UEi TEST INSTRUMENTS www.ueitest.com Tel: 1-800-547-5740 CO CO-H2 CO 2 NO Serial No. N. O2 O. Sensors Fitted: C. D. E. H. O2/Eff CO/CO2 Temp I. Exch Test Aux Status P. ROOM CO F. J. MP PRS v3 TE LINK K. CO 2 CO C165+ Com busti G. o n An er alyz L. M. Q. [Top View] S. R. T2 P2 T1 P1 T. U.
A. On/ Off (Power) Button B. 6 Line Backlit Display • Press any button to turn Back light on (backlight duration can be set in Menu/ Screen/Backlight) C. UP/DOWN Button • Short press to navigate “UP” or “DOWN” D. Pump Toggle button: Long press to pump on/off, short press holds the readings and the screen flashes E. ENTER Button • Short press select current option displayed F. Rotary Dial G. Protective Rubber Boot With Magnets H.
MENU SCREEN - C165+ START UP Press ”Power” Button for 2 seconds. The analyzer will perform a 10 second purge. This should be done in fresh air. Rotate Selector Dial to MENU to set up or customize your settings. Selected parameter are highlighted with Display Line Lights on either side. and Use options. ARROW BUTTONS to scroll menu O2/Eff Press Button to edit a parameter. CO/CO2 Temp Exch Test Aux Status ROOM CO Press field. MP TE ARROW BUTTONS to change CO 2 Press and contents of field.
BEFORE TESTING CHECKLIST o Clean particle filter o Water trap and probe line are free of condensate o All hose and thermocouple connections are properly secured o Water trap is fitted correctly o Power on and zero o Flue gas probe is sampling ambient FRESH air Remove the rubber boot and red drain plug. Allow the water to drain out. Re-insert the rubber plug and replace boot cover. Remove protective rubber boot. Remove water trap from analyzer.
BASIC CO/COMBUSTION ANALYSIS T2 P2 T1 P1 O2/Eff CO/CO2 O2/Eff Temp Status ROOM CO CO/CO2 Temp Exch Test Aux Exch Test Aux Status ROOM CO v3 PRS MP LINK CO 2 CO C165+ Com busti Insert Flue probe into stack. Adjust cone so end of the probe is approximately at the center of the stack. NOTE: You will have to drill a hole at least 3/8”. Use appropriate sealing method after testing.
PRESSURE TEST T2 P2 O2/Eff CO/CO2 T1 P1 Temp Exch Test Aux Status ROOM CO MP PRS v3 TE LINK CO 2 CO C165+ Com busti on A ze n aly r Rotate the selector dial to PRS/TEMP. Press to Zero Pressure sensor. No connections when zeroing pressure. Connect true draft hose to P1. Connect second probe to P2 for differential pressure. Place combustion probe or true draft probe tip in the flue to measure draft. Connect static pressure hose to measure differential pressure.
HEAT EXCHANGER TEST O2/Eff CO/CO2 Temp Exch Test Aux Status ROOM CO MP PRS v3 TE LINK CO 2 CO C165+ Com busti yze o n A n al r Rotate dial to EXCH TEST. Press to start. or Scroll FUELTYPE. Press to select Call for heat on the system. Observe and wait for O2 readings to stabilize.
ROOM CO TEST CO/CO2 O2/Eff Temp Exch Test Aux Status ROOM CO MP PRS v3 TE LINK CO 2 CO C165+ Com busti o n A n al yze r No probes or hose connections required for this test. Rotate dial to ROOM CO. Press to select GENERAL test type. CO reading will be logged every minute for 30 minutes. Test results and LOG number will display when done. Readings will be stored once the 30 minute test is completed. Press to print the results.
SETUP USER FUEL O2/Eff CO/CO2 Temp Exch Test Aux Status ROOM CO MP PRS v3 TE LINK CO 2 CO C165+ Com busti o n A n al yze r Rotate dial to Menu UNITS shows press proceed. to POST TESTING POWERING OFF When you power off the C165+, there is a 10 second purge. CAUTION Remove the probe from the flue and allow analyzer to purge with fresh air until readings return to zero. O2 to 20.9% CAUTION: probe tip will be HOT. Make sure you do not exceed the analyzer’s operating specifications.
WARRANTY The C165+ combustion analyzer is warranted for one year. Warranty is a manufacturing defect identified within the product’s warranty period. Warranty does not cover batteries, fuses, test leads, probes, consumables physical damage and misuse. For full warranty details please refer to the Owner’s Manual on the C165+ product page (see the link above). If a UEi product fails under warranty, we will replace it but we need the unit back so we can identify and fix the problem.
ANNUAL RECERTIFICATION SERVICES Complete care for your C160 series combustion analyzer. When you: BACKED BY Request Recertification or Service Online Within 1 Year of Purchase or Service UEi will: 6-Year Warranty: All UEi combustion analyzers have a standard 1-year warranty. Each recertification includes a 30-point inspection and extends the warranty for 1 more year for up to 6 years from the date of warranty. Contractors who book recertification of a C160 series analyzer at www.ueitest.