Temperature Measurement
Parameter Range Resolution Accuracy
Flue Temperature 32˚ to 1112˚F
(0˚ to 600˚)
0.1˚F (0.1˚C) ±0.5˚F (0.5˚C)
Inlet Temperature
(internal Sensor)
32˚ to 1112˚F
(0˚ to 600˚)
0.1˚F (0.1˚C) ±1˚F (1˚C)
Inlet Temperature
(External Sensor)
32˚ to 1112˚F
(0˚ to 600˚)
0.1˚F (0.1˚C) ±0.5˚F (0.5˚C)
Flue Gas Measurement
Parameter Range Resolution Accuracy
Carbon Monoxide 0 to 2000 ppm 1 ppm ±3 ppm or ±5% rdg
(whichever is greater)
Carbon Dioxide 0 to 20% 0.1% ±0.3% volume
Calculations
Parameter Range Resolution Accuracy
Oxygen 0 to 21% 0.1% ±0.3% volume
Efficiency (Net or
Gross)
0 to 99.9% 0.1% ±1% rdg
Efficiency High (C) 0 to 119.9% 0.1% ±1% rdg
Excess Air 0 to 119.9% 0.1% ±0.2% rdg
CO/CO
2
Ratio 0 to 0.9999 0.0001 ±5% rdg
Range Resolution Accuracy
Pressure (Differential) ±80 mBar 0.1 mBar ±0.5% FSD
Specifications
Preprogrammed Fuels
Pellets, Light Oil, LPG, Butane, Propane, Natural
Gas, Bio Oil, Heavy Oil
Battery Life
>8 hours (continuous with pump on)
Certification
The C160 Series is TUV-tested and certified
to EN 50379, Parts 1-3 in accordance to 1st
German Federal Emission Control Ordinance
(BimSchV)
Operating Conditions
Temperature 32˚ to 113˚F (0˚ to 45˚C)
Humidity 15% to 90% R.H.
Power Supply Rechargeable Batteries, USB Charging
Dimensions 8.54" x 4.18" x 1.86"
Weight 1 lb., 5.1 oz.
No O2 Sensor - What it means to you
Gas Flow
CO
Sensor
CO2
CO2
Measured CO & CO2
Calculated O2
Replace the Oxygen sensor
with UEi’s “EOS” Carbon
Dioxide (CO2) sensor.
Our technology protects
the sensor because flue
gases don’t make direct
contact with it.
Because the sensor is not
wearing down, there’s less
risk of inaccurate calculations.
No O2 sensor to replace
means lower cost of ownership.
The Solution
Direct contact between
flue gases and O2 sensors
will eventually wear down
the sensor and force costly
replacement.
Even when your analyzer
is “off” the O2 sensor is
still in Oxygen, accelerating
the process.
Over your analyzer’s
lifetime, the cost adds up.
The Challenge
Gas Flow
CO
Sensor
O2
Sensor
Measured CO & O2
Calculated CO2
One Less Sensor to Worry About!
Backed by Industry-Leading Service
2-Business day turnaround on standard recertification
Flat rate pricing
ISO/IEC Accredited Facilities (Indianapolis and Vancouver B.C.)
NIST and ISO Accredited Service