Manual

diSPlay ParameterS oN uei CombuStioN aNalyzerS
• TF(FlueTemperature)UsedtocalculatenettemperatureinFahrenheitorCentigrade.Will
show ambient temperature after fresh air calibration and- O C - ’iftheflueprobeisdisconnected.
• T (Net Temperature)Indicated by T (Greek letter delta for difference)
- Differential of flue temperature minus ambient (or inlet if set) temperature
-DifferentialofT1-T2onC125andC127
• O2 (Oxygen) Direct reading of oxygen from the O2 sensor reading in percentage (%)
• COCarbonMonoxidereadingdisplayedinppm(partspermillion).‘----’or‘-OC-’is
displayed if there is a fault with the CO sensor or the instrument has not set to zero correctly, switch
off instrument and try again.
• CO2 Calculated from the fuel selected and the measured O2 level
- Will display “----“ or “-O>-” if the oxygen level is too high to calculate CO2
-Onlydisplaysareadingduringcombustiontest.‘----’or‘-O>-’isdisplayedwhileinfreshair.
X-Air(ExcessAir)
- Indicated by _(Greeksmallcaselambda)or‘X’onscreen
- Indicated by XAIR % on the printout
-Excessairisacalculationofthepercentageofoxygenabovewhatistheoreticallyrequiredfor
complete combustion
- Isrequiredtocompletelyburnallofthefuelduetopoormixingandtoassistinventingfluegases
- Onlydisplaysareadingduringcombustiontest.‘----’or‘-O>-’isdisplayedwhileinfreshair.
• Eff (Efficiency)
- Indicated by _ (Greek letter eta) followed by G for gross, N for net or C for condensing on the
screen if the analyzer has these options
-IndicatedbyEff(G)ontheprintoutifsettoGross(usedintheStates)
-Eff(N)orEff(C)fortheotherscales(basedonstandardsforthatlocation)
- Is calculated based on gas readings, net temperature and fuel selected.
- This value is combustion efficiency, not appliance efficiency
• Loss (in the display)
- Total losses calculated from Combustion Theory. This is the summation of the next three param-
eters
Dry%:CalculatedheatlostturningtheCarboninthefueltoCarbonDioxide(CO2)
•Wet%:CalculatedheatlostturningtheHydrogeninthefuelintowater(H20)
COLoss%:CalculatedlossduetopartiallyburntCarbon.AnyCarbonMonoxide
(CO)in the flue has the potential to be turned into Carbon Dioxide and release more
heat, thus this heat is lost up the flue
• CO AF(CarbonMonoxideAir-Free)Readingreferencedtoanoxygenlevelof0%.Donotconfusethis
readingwiththeactualCOreadingasdetailedabove.SeetheCombustionEfficiencyCalculation
sections for more details.
AMB AirINLETtemperatureusedtocalculatetheNETtemperature.
 • CO/CO2 (CO/CO2 Ratio) The ratio of measured CO divided by calculated CO2.
-Itgivesanindicationof:
•Howgoodagassampletheinstrumentisreading.
Howcleantheboilerisrunning.
-Forexample:Aneworcleandomesticboilerwilldisplayaratiooflessthan0.004,aunitinneed
ofcleaning0.004-0.008andaunitinneedofmajoroverhaulwillshowgreaterthan0.008.
-Onlydisplaysareadingduringcombustiontest.‘----’isdisplayedwhileinfreshair.