Operation Manual

english
E
EN-48 EN-49
english
E
Technical informationTechnical information
Pressure display range: 0-300 mmHg
Measurement range: Systolic: 50-250 mmHg
Diastolic: 40-180 mmHg
Pulse measurement: 40-160 pulses/min
Error limits: Blood pressure measurement: corresponds to EN
1060 Part 3
Pressure measurement: ± 3 mmHg
Pulse measurement: ± 5%
Serial number: The unit bears a serial number SN which provides
clear identication.
Power supply: Battery type: 2 x 1.5 V, AAA, LR 03, durability: over
700 measurements in 2 years
Cuff: Cuff type visomat H5, wrist circumference 12.5 cm -
21.5 cm
Operating conditions: Ambient temperature 10 to 40 °C
Relative humidity up to 85%, non-condensing
Storage and
transportation
conditions:
Ambient temperature -5 to 50 °C, Rel. air humidity
up to 85% not condensing,
Air pressure 700 to 1040 hPa
Deation valve: Electronically controlled
Automatic switch-off: approx. 3 minutes after end of measurement
Protection class IP: IP 20: Protected against solid foreign particles with
a diameter of more than 12.5 mm, no protection
against water.
Expected service life: 5 years
Classication: Internally powerd ME Equipment
2. Customer service
The unit may only be repaired by the manufacturer or by an expressly authorised
organisation.
Please contact:
UEBE Medical GmbH
Zum Ottersberg 9
97877 Wertheim, Germany
Phone: +49 (0) 9342/924040
Fax: +49 (0) 9342/924080
E-mail: info@uebe.com
Internet: www.uebe.com
Unit type: Digital automatic unit with electric pump for measuring
blood pressure at the wrist
Dimensions: approx. W = 70 mm x H = 70 mm x D = 80 mm with
cuff
Weight: 121 g with cuff, not including batteries
Display: LCD display (liquid crystal display) for measured values
and check displays
Memory: 60 measured values (stored automatically) and average
value (A)
Measurement
procedure:
Oscillometric determination of systole, diastole and pulse
Reference procedure
of clinical testing:
Auscultatory measurement
Inating pressure: Automatically controlled
3. Technical data