Operation Manual

English
English
EN-56 EN-57
Technical informationTechnical information
EE
1. Error messages
Error Possible cause Remedy
Display Err - 300 Excessive cu
pressure. The cu
was inated to its
maximum due to the
movement of the
arm or body.
The air hose may be
kinked or the air plug
blocked.
- Repeat the measurement
- Do not move your arm
- Do not speak
Inspect the hose line and air
plug for blockages
Display Err - 1 Measurement error;
measurement could
not be carried out
due to movement or
speaking.
- Repeat the measurement
- Do not move your arm
- Do not speak
The acquired values
lie outside the mea-
suring range of the
device (see „Technical
data“ on page 58).
Repeat the measurement
Display Err - 2 Ination error;
pressure built up too
quickly or too slowly.
Check that the air connector
is tted properly, repeat
measurement.
Fault has been iden-
tied. The cu is not
correctly attached
- Repeat the measurement
- Do not move your arm
- Do not speak
Display Err - 3 Deation rate too
high or too low.
Call the Service hotline if
this occurs repeatedly.
Display 0
0
Start/Stop button
was accidentally
pressed while the
batteries were being
replaced or while the
mains adapter was
being plugged in.
Switch unit o and then on
again using the Start/Stop
button.
Error Possible cause Remedy
Arrhythmia, mo-
vement disorders,
trembling, wobbling
(objects), deep bre-
athing etc.
Repeat measurement after
resting for 3-5 minutes.
Listen out for the device‘s
pulse beep signals. Possible
inuences on the measure-
ment results (depending on
the severity of the arrhyth-
mia) must be discussed with
your doctor!
The motion control
detected movement
during the measure-
ment.
- Repeat the measurement
- Do not move your arm
- Do not speak
Measured values/
pulse values are
extremely high or
low.
Too little rest before
taking the measure-
ment. Movement or
talking during the
measurement.
Repeat the measurement
after a break of 3-5 minutes.
Pay attention to the usage
information on page 43.
Incorrect measure-
ment posture
Repeat the measurement
and maintain the correct
posture. (Page 47)
Measurements
taken in direct
succession
produce dierent
measured values.
Blood pressure is not
a xed value. It can
go up or down by
more than 20mmHg
within a few minutes.
In order to obtain compara-
ble results, always carry out
the measurements under
the same conditions.
Measurement was
interrupted, the
cu deated to 0
and the measure-
ment restarted.
Device detects a fault
or that the diastolic
value is too low.
Device shows no fault.
Repeat measurement
Movement during
measurement
If necessary, interrupt mea-
surement and repeat after 5
minutes‘ rest.
No display after
device is switched
on
Batteries inserted
incorrectly
Check position of batteries.
Batteries at Replace batteries.
Battery contacts
dirty.
Clean battery contacts with
a dry cloth.
Dashes in display
after pressing
Memory button
No measured values
have been stored.
Repeat measurement