User's Manual
g. Adjust the SCREEN fine control until the SCREEN VOLTS
meter indicates the voltage specified on the tube test data roll chart
(par. 12).
h. Set the BIAS fine control to its extreme clockwise (maxi-
mum) position.
i. With the PRESS TO TEST P5 switch held in its operated
position, gradually reduce the bias voltage by adjusting the BIAS
fine control until the tube strikes its firing potential. The firing
potential is indicated by a sudden increase in load current shown
on the PLATE meter and by a sudden decrease on the PERCENT
QUALITY meter.
Note. The tube will not conduct (as indicated by an increase in load current
shown by the PLATE meter) until the critical grid-bias voltage is reached.
For example, for a type 2D21W or type 2050W tube, the bias voltage at which
conduction occurs should be between 1.5 and 3.0 volts.
j. The voltage at which the tube strikes is read on the GRID
BIAS VOLTS meter. Record the grid voltage at the instant and
at the point that the pointer starts to deflect. This is the striking
voltage of the tube under test. Repeat the test to confirm the read-
ing. Minimum limits for striking voltages of thyratrons are given
on the tube test data roll chart.
k. When the test is completed, set the ON-OFF power switch
to the OFF position, remove the tube under test from the test
socket, and return all switches and controls to their safety posi-
tions (par. 15d).
27. Electron-Ray Indicator Test
There are two types of electron-ray (ER) indicator tubes that
can be tested with the TV–2(*)/U, Type 6U5 and similar tubes
have a triode section and a single indicator (one shadow). Type
6AF6 and similar tubes have no triode section, but have twin
indicators (dual shadow).
Follow the procedure below to check
the opening and closing action of the eyes only. Indicators with
triode sections also must be tested for emission (i below).
a. Perform the operations indicated in paragraphs 15 and 16.
b. Operate the FUNCTION switch to the TEST position.
c, Operate the PLATE-SCREEN RANGE switch to the posi-
tion specified on the tube test data roll chart.
d. Adjust the PLATE and SCREEN fine controls until the
exact voltages specified on the tube test data roll chart are indi-
cated on the corresponding meters.
e. Depress the PRESS TO TEST P4 switch. If the tube is a
single indicator, the indicator should be closed; if the tube is a
twin-indicator tube, such as the 6AD6 or 6AF6, indicator No. 1
should be closed and indicator No. 2 should be open.
31