User's Manual

RANGE switch are set as in a(2) through (4) above, and the
FILAMENT VOLTS meter indicates 6.3 volts,
(1) Set the PLATE switch to the 5 position.
(2) Set the PLATE-SCREEN RANGE switch to the S posi-
tion.
(3) Adjust the PLATE fine control until the PLATE meter
indicates 20 volts ac.
(4) Turn the quality SHUNT control to 90.
(5) Set the GM-SIGNAL RANGE switch to the A position.
(6) For the TV-2/U, aline the dot on the knob of the GM
CENTERING control with the dot on the panel. For
the TV-2A/U and TV–2B/U, aline the line marking on
the GM CENTERING control with the line marking on
the panel.
(7) Depress the PRESS TO TEST P2 switch and read the
PERCENT QUALITY meter. If the indication is less
than the minimum limit of 25, discard the tube. If the
indication is satisfactory, measure the emission of diode
No. 2.
h. Measurement of Emission (Diode No. 2). To measure the
emission of diode No. 2, adjust the tube test data roll chart so
the third line on which the 6SQ7 tube appears is between the
horizontal red lines. The FUNCTION switch, all SELECTORS
switches except the PLATE switch are set as in g above, and the
FILAMENT VOLTS meter indicates 6.3 volts.
(1) Set the PLATE switch to the 4 position.
(2) With the PLATE-SCREEN RANGE switch on S, ad-
just the PLATE fine control until the PLATE meter
indicates 20 volts ac.
(3) The quality SHUNT control remains on 90.
(4) The GM-SIGNAL RANGE switch remains on S.
(5) For the TV-2/U, aline the dot on the knob of the GM
CENTERING control with the dot on the panel. For
the TV-2A/U and TV-2B/U, aline the line marking on
the GM CENTERING control with the line marking on
the panel.
(6) Depress the PRESS TO TEST P2 switch and read the
PERCENT QUALITY meter. If the indication is less
than the minimum of 25, discard the tube. If the indica-
tion is satisfactory, the tube is good.
36