Safety Instructions

Display Description Action
An error message indicates a problem with
the meter.
Remove batteries and wait for 1 minute. Reinstall
batteries to see if meter is working properly. Please
contact customer service if the problem persists.
An error message indicates a problem with
the test strip.
Review the instructions and try again with a new test strip.
Please contact customer service if the problem persists.
An error message indicates a problem with
the meter.
Remove batteries and wait for 1 minute. Reinstall
batteries to see if meter is working properly. Please
contact customer service if the problem persists.
An error message indicates insucient
volume of blood sample for the test.
Follow the instructions and try again with a new test strip,
then applying a sucient blood sample to the test strip.
Please contact customer service if the problem persists.