Safety Instructions
1. Battery power is low, icon ashes for 2
seconds. Meter can work normally but will only
provide approximately 10 more measurements.
2. Battery is depleted, icon ashes continuously.
Meter will be shutdown in 5 seconds.
Display Description Action
The preset reminder alarm is ringing Press any button to turn o the alarm or wait for the alarm to
be turned o automatically in 30 seconds.
Replace with 2 AAA batteries.
1. An error message indicates that the
hematocrit of the blood sample is too
low for the system to perform a test.
2. Use control solution to check the meter
and test strips.
1. Follow the instructions and try again with a new test strip.
Please contact customer service if the problem persists.
2. Follow the instructions to enter control solution mode,
and try again with a new test strip.
An error message indicates that the
hematocrit of the hematocrit of the
blood sample is too high for the system
to perform a test.
Follow the instructions and try again with a new test strip.
Please contact customer service if the problem persists.